Increased False Positives vs Increased Interference Flags: Comparison of 2 Ethyl Glucuronide Immunoassays.
Alexandria N Jackson, Sara P Wyness, Sonia L La'ulu, Jessica M Boyd, Kamisha L Johnson-Davis
ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT, United States. · Department of Pathology, University of Utah Health, Salt Lake City, UT, United States.
Key points
From the paper's conclusion- Selecting an assay for the laboratory requires understanding the strengths and limitations of the method. For our work flow, our lab chose the assay with more interferences, since our work flow reflexes presumptive positive immunoassay results and interferences to confirmation by mass spectrometry. For labs that use an immunoassay screen-only approach, selecting an immunoassay that generates a higher rate of interference flags could make result interpretation difficult for the client and raise concerns for specimen adulteration.
Abstract
BACKGROUND: Screening for alcohol use in a population to treat pain and substance use disorders is important for monitoring adherence to treatment plans. Urine screening for ethanol exposure by immunoassay can provide a cost-effective testing option, with a faster turnaround time to result in comparison to LC-MS/MS. The purpose of this study was to compare two ethyl glucuronide (EtG) immunoassays to determine which method would be better suited for our laboratory work flow and patient population. METHODS: Two EtG immunoassays were evaluated to assess false-positive results and interference flags for alcohol biomarker testing. The EtG enzyme immunoassays were verified and put into use for patient testing during 2 separate timeframes. RESULTS: Both assays compared qualitatively well to LC-MS/MS for positive EtG (>500 ng/mL) patient samples. In our patient population, the ARK EtG assay had a presumptive positivity rate of 13.81%, a false-positive rate of 1.35%, and an interference rate of 0.45%. The Immunalysis assay had a 9.87% presumptive positivity rate, a false-positive rate of 0.15%, and a 2.86% interference rate. CONCLUSIONS: Selecting an assay for the laboratory requires understanding the strengths and limitations of the method. For our work flow, our lab chose the assay with more interferences, since our work flow reflexes presumptive positive immunoassay results and interferences to confirmation by mass spectrometry. For labs that use an immunoassay screen-only approach, selecting an immunoassay that generates a higher rate of interference flags could make result interpretation difficult for the client and raise concerns for specimen adulteration.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- False-Positive Urine Ethyl Glucuronide Results in Pregnancy Caused by Escherichia coli Urinary Tract Infection.O&G open · 2026
- Hours Since Last Drink Is Associated With Multiple Urine Ethyl Glucuronide Cutoffs Among Individuals With Alcohol Use Disorder.Drug testing and analysis · 2026
- 49 Assay Performance and Test Utilization: A Comparison of E-Z Cup Point-of-care Device and Alinity c Automated Urine Drug Screening PlatformsAmerican Journal of Clinical Pathology · 2026
- Improved benzodiazepine detection with a beta-glucuronidase-containing immunoassay and implications on laboratory workflow.Clinica chimica acta; international journal of clinical chemistry · 2026
- Analytical evaluation of a urine immunoassay for 2-ethylidene-1,5-dimethyl-3,3-diphenyl pyrrolidine (EDDP) with focus on cutoff optimization.Journal of forensic sciences · 2026
- Rapid screening of ethylene glycol and diethylene glycol in raw materials and medicinal syrups using low-cost field deployable assaysScientific Reports · 2025