Improved benzodiazepine detection with a beta-glucuronidase-containing immunoassay and implications on laboratory workflow.
Samuel A Krug, Abdol Aziz Ould Ismail, K Aaron Geno, Lynn A Brunelle
Department of Pathology and Laboratory Medicine, Dartmouth-Hitchcock Medical Center, Lebanon, NH, United States; The Geisel School of Medicine at Dartmouth, Hanover, NH, United States. · Department of Pathology and Laboratory Medicine, Dartmouth-Hitchcock Medical Center, Lebanon, NH, United States.
Key points
From the paper's conclusion- The frequency of urine samples sent for confirmatory benzodiazepine testing doubled after implementing BNZ2. For benzodiazepines with a known glucuronide metabolite, we observed an increase in screen positivity in the outpatient setting compared to the original BENZ Plus assay.
Abstract
INTRODUCTION: Benzodiazepines (BZDs) are widely-prescribed sedatives that act on GABAA receptors but are also frequently misused, prompting routine urine drug screening. Traditional immunoassays, like the Roche BENZ Plus assay, have limited cross-reactivity with phase II BZD metabolites and certain novel psychoactive substances (NPSs), leading to false negatives. This study evaluated the impact of implementing the second-generation Roche BNZ2 assay, which includes β-glucuronidase, to improve detection of glucuronide metabolites. MATERIALS AND METHODS: Confirmatory results from screen-positive urine samples were obtained and analyzed for the 8-month periods before and following laboratory conversion to the updated screen. Additionally, analytical standards for bromazolam, etizolam, flubromazolam, alprazolam, hydroxyalprazolam, zolpidem, lorazepam, and lorazepam glucuronide were purchased and spiked into drug free urine for cross-reactivity studies. RESULTS: We observed an increase from 5.2% presumptive positive samples needing confirmation after analysis with the BENZ Plus assay compared to 10.0% needing confirmation after analysis with the BNZ2 assay. The three most prevalent benzodiazepines in screen-positive inpatients were lorazepam, midazolam, and α-hydroxymidazolam, regardless of the assay. We observed an increase in prevalence of lorazepam, oxazepam, 7-aminoclonazepam, and zolpidem phenyl-4-carboxylic acid for outpatient samples analyzed with BNZ2 compared to BENZ Plus. Cross-reactivity studies using spiked samples showed generally good agreement with manufacturer data across structural benzodiazepine analogs. CONCLUSIONS: The frequency of urine samples sent for confirmatory benzodiazepine testing doubled after implementing BNZ2. For benzodiazepines with a known glucuronide metabolite, we observed an increase in screen positivity in the outpatient setting compared to the original BENZ Plus assay.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 1 Expanding Benzodiazepine LC-MS/MS Confirmation Improves Diagnostic Accuracy and Clinical Interpretation in Toxicology TestingAmerican Journal of Clinical Pathology · 2026
- B-381 Comparison of Benzodiazepines Positivity Rates in Serum and Urine Specimens by LC-MS/MS in a Clinical Laboratory SettingClinical Chemistry · 2026
- Standardization of a Protocol to Quantify Benzodiazepines in Human Urine Samples Employing Gas Chromatography-Mass SpectrometryA Green Analytical Chemistry Approach.ACS omega · 2026
- In vitro γ-aminobutyric acid A (GABAA) receptor activity and binding interactions at the α+/γ2– interface of 53 prescription and designer benzodiazepinesCommunications Chemistry · 2026
- Designer Benzodiazepines in Norwegian Police Casework: Blood Concentrations Related to Clinical Impairment.Journal of analytical toxicology · 2026
- Comparison of Immunoassay Results near Cutoff Values with LC-MS/MS in Individuals with Substance Use Disorder Under Probation.Metabolites · 2026