Ten years follow-up data on a leadless pacing system – real life experience in comparison to data of the prospective trial
S Winter, Q Nguyen, T Tuscher, s Fischer, J M Sinning
Key points
From the paper's conclusion- We were also able to show stable long-term system parameters in the clinical setting of up to ten years of follow up with a very good battery longevity of about 12,0 years.
Abstract
Introduction The Micra™ leadless intracardiac pacing system has been introduced and implemented into clinical routine in 2015, with its next generation MICRA AV ™ in 2020. Feasibility, implantation safety, and acute success have been proven in the setting of controlled studies. Even apart from controlled studies, this leadless pacemaker technology iss beneficial for the individual patient. We aim to report our single center follow-up (FU) data with a follow-up of up to sten years. Methods In 412 patients (188 men; age: 81±5,7 y), Micra™ implantation was performed. Pacemaker interrogation was performed one to seven days after implantation and during FU (1; 3; 6, then every 6 month) up to 120 months. Data were assessed in a real-life setting and compared with existing data of a controlled prospective trial. Results The implantation was successful in all 412 attempts with only two pericardial effusion as major complication, which needed pericardiocentesis, two complications with the groin which required operation. During follow up there were three patients developing severe heart failure symptoms resulting in an implantation of a CRT-device and switching of the leadless pacemaker system. One device had to be changed because of early unacceptable increase in threshold. The average acute thresholds, sensing and impedance after system release were: 0.56±0.36V@0.24ms; 10,1 ± 6,14mV and 754±1 196 Ohm. During follow up of up to 10 years neither, pacemaker failure, nor infections were reported. Measurements were reevaluated for long-term thresholds, sensing and impedance: 0.75 ± 0.38V @ 0.24 ms; 18,38 ±2,75 mV and 525 ± 102 Ohm. During 10 years, in average no significant changes from acute to long-term measurements were detectable. In comparison to the controlled trial, our measurements of this real-life cohort were very similar. Conclusion In a real-life setting, the implantation of the leadless Micra™ system demonstrates a high rate of implantation success without major complications. We were also able to show stable long-term system parameters in the clinical setting of up to ten years of follow up with a very good battery longevity of about 12,0 years. Treshold Study vs. Paper
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A decade of leadless pacemaker implantation: technical performance, safety and device longevityEuropean Heart Journal · 2026
- Pacing‐Induced Cardiomyopathy in Leadless and Transvenous Pacemakers: A Meta‐Analysis and Meta‐RegressionPacing and Clinical Electrophysiology · 2026
- Infection and Retrieval of Leadless Pacemakers: Current Challenges and Future DirectionsJournal of Clinical Medicine · 2026
- Adverse Event Profiles of Leadless Pacemaker Systems Reported to the FDA MAUDE Database (2021-2026).Journal of cardiovascular electrophysiology · 2026
- MicraAV Versus Dual-Chamber Transvenous Pacemakers in Patients with Atrioventricular Block Following Transcatheter Aortic Valve Replacement: A 12-Month Single-Center Retrospective Study.Journal of clinical medicine · 2026
- Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohortEP Europace · 2025