Cardiopulmonary Exercise Testing in Patients with Cardiac Pacemakers: From Device Optimization to Individualized Cardiac Rehabilitation.
Bogdan Caloian, Carmen Silvia Caloian, Raluca Tomoaia, Diana Andrada Irimie, Florina Iulia Fringu, Dan Horatiu Comsa, Gabriel Laurentiu Cismaru, Gabriel Nicolae Gusetu,
Department of Cardiology-Rehabilitation, Faculty of Medicine, University of Medicine and Pharmacy "Iuliu Hatieganu", 400347 Cluj-Napoca, Romania. · Department of Periodontology, Faculty of Dental Medicine, University of Medicine and Pharmacy "Iuliu Hatieganu", 400347 Cluj-Napoca, Romania.
Key points
From the paper's conclusion- Reprogramming alters the result of the test within 24 h, so the settings should be corrected before rehabilitation begins.
Abstract
Permanent cardiac pacing abolishes symptomatic bradycardia, yet a substantial proportion of recipients remain limited on exertion. Chronotropic incompetence is present in about half of pacemaker-wearing patients when assessed by the Wilkoff index, and the rate-responsive function intended to correct it is frequently left at the settings applied at implantation. Read across the whole cardiopulmonary exercise test rather than at peak effort, the relationship between heart rate and oxygen consumption distinguishes an inadequate sensor slope, an activity threshold set too high, a reaction time out of step with the onset of exercise, and an upper rate limit reached prematurely, each with a distinct correction. Reprogramming alters the result of the test within 24 h, so the settings should be corrected before rehabilitation begins. The objective is not the highest achievable rate but the individual ceiling above which the ventricle no longer converts frequency into output. Because heart rate is partly device-determined, training intensity should be anchored to the ventilatory threshold, to the corresponding workload and to perceived exertion, with heart rate being monitored for safety rather than used to set intensity. The aim of this review is to describe how cardiopulmonary exercise testing guides device reprogramming and training prescription in the pacemaker-wearing patient.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rate response of temperature-based sensors in leadless pacemakers during exercise test: a small cohort studyEuropean Heart Journal · 2026
- Accelerated pacing rates to improve exercise tolerance, quality of life, and arrhythmia burden in patients with evidence of heart failure with preserved ejection fraction: The PACE-UP rationale and design.Heart rhythm O2 · 2026
- Diagnosing chronotropic incompetence: a scoping review of current approaches and their application in a cardiac rehabilitation cohortEuropean Journal of Preventive Cardiology · 2026
- Cardiopulmonary exercise performance in physiologic versus conventional pacingEuropean Heart Journal · 2026
- Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery.JACC. Case reports · 2026
- Force–frequency–guided pacing improves cardiac output and attenuates severe functional tricuspid regurgitationEuropean Heart Journal · 2026