Comparison of VASCADE MVP XL and Figure of 8 suture for vascular closure of the Faradrive sheath for pulsed field ablation
T Tanko, J Ahn
Key points
From the paper's conclusion- Conclusion In this retrospective single center study, using VASCADE MVP XL for closure of the Faradrive delivery sheath is comparable to figure of eight closure with high deployment success and low risk of bleeding and vascular complication.
Abstract
Background Hemostasis following atrial fibrillation (AF) ablation is often achieved with figure of 8 suture (F8) and vascular closure devices. The VASCADE MVP XL closure device is indicated for closure of large bore sheaths to a 15Fr outer and 12Fr inner max diameter. Farapulse pulsed field ablation (PFA) is performed through the Faradrive sheath with a 16.8 Fr outer and 13 Fr inner diameter. Purpose The purpose of the study is to evaluate the efficacy and safety of VASCADE MVP XL compared with figure of eight stitch (F8) closure of the Faradrive delivery sheath in patients undergoing PFA ablation and same day discharge. Methods This was a single center retrospective study, involving 118 patients who underwent Farapulse PFA ablation over a 6 month period and closed with either the VASCADE MVP XL or F8. All patients underwent reversal with protamine prior to sheath removal. The F8 stitch was removed prior to discharge. Outcomes were assessed immediately post procedure and for the duration of observation prior to discharge. Results Over a 6-month period, patients were selected retrospectively. A total of 118 patients underwent outpatient PFA for AF. 59 underwent VASCADE MVP XL closure device and 59 underwent FO8 closure. There was no significant difference in Protamine dose between groups. In the VASCADE MVP XL arm(N=59): Success of deployment/closure- 59(100%), bedrest time(avg)-3hours 59(100%), minor bleeding- 0(0%), major bleeding- 0(0%), vascular complication- 0(0%), same day discharge- 59(100%). In the F8 arm(N=59): Success of deployment/closure- 58(98.3%), bedrest time(avg)- 3hours 57(96.6%), 4hours 2(3.4%), minor bleeding- 2(3.4%), major bleeding- 1(1.7%), vascular complication- 1(1.7%), same day discharge- 57(96.6%). Minor bleeding - bleeding that required additional manual pressure or small hematoma. Major bleeding - bleeding associated with >2 g/dl drop in hemoglobin, requiring transfusion or large hematoma. Vascular complication - deep vein thrombosis, arteriovenous fistular, pseudoaneurysm or requiring surgical repair. Conclusion In this retrospective single center study, using VASCADE MVP XL for closure of the Faradrive delivery sheath is comparable to figure of eight closure with high deployment success and low risk of bleeding and vascular complication.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety and Efficacy of Two Venous Closure Devices Compared With Manual Compression After Atrial Fibrillation Ablation.Journal of arrhythmia · 2026
- A dual-phase study of a bioabsorbable venous vascular closure system: identifying predictors of failure and ensuring long-term safety in over 1,000 proceduresEuropean Heart Journal · 2026
- Venous Vascular Closure System Versus Conventional Figure of Eight Suture Following Single Shot Device Pulmonary Vein Isolation: The VASC-AF Study.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- Safety and Efficacy of the VASCADE MVP® XL Venous Vascular Closure System for Management of Femoral Venotomy Following Catheter‐Based Interventions Utilizing 16‐17F OD Sheaths: The AMBULATE EXPAND TrialJournal of Cardiovascular Electrophysiology · 2026
- Vascular closure with a suture retention device versus suture-mediated closure system after large bore venous access for electrophysiologic procedures.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Device- and patient-related complications of pulsed-field ablation catheters: a FDA MAUDE database analysis comparing FARAWAVE and VARIPULSEEuropean Heart Journal · 2026