Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing

Vascular closure with a suture retention device versus suture-mediated closure system after large bore venous access for electrophysiologic procedures.

Aashish Katapadi, Jiaqi Mi, Eli Herink, Michael Klotz, Ahmed El-Ballat, Karnik Patel, Rajesh Kabra, Peter Park,

HCA Midwest Health, Overland Park, KS, USA. · Kansas City Heart Rhythm Institute, 5100 W 110th Street, Suite 200 Overland Park, Overland Park, KS, 66211, USA.

Observational

Key points

From the paper's conclusion
  • The two devices have similar hemostasis outcomes for LBVA following LAAO, but each has unique considerations. LockeT was associated with fewer access-site complications and device failures; Perclose was associated with faster TTA. Nevertheless, additional studies are still needed.

Abstract

BACKGROUND: The LockeT is a suture retention device that may be a postoperative closure option for electrophysiologic procedures requiring large-bore vascular access (LBVA) but warrants further evaluation. OBJECTIVES: We compare the use of LockeT to a suture-mediated vascular closure device for LBVA after left atrial appendage occlusion (LAAO). METHODS: We performed a retrospective study of LBVA after LAAO from March 2023 to August 2024 using LockeT or Perclose ProGlide. Endpoints included access-site-related complications, time to hemostasis (TTH), time to ambulation (TTA), and same-day discharge (SDD). RESULTS: We identified 208 patients (n) who underwent either LockeT (n = 86; N = 172) or Perclose (n = 122; N = 362) devices. Even after propensity score matching (LockeT: n = 86; and Perclose: n = 86), those undergoing LockeT had few differences in baseline characteristics and were associated with more frequent Amulet (62.8% vs. 16.3%) than Watchman (37.2% vs. 68.6%) LAAO implantation. The LockeT was associated with no difference in TTH (20 [interquartile range (IQR) 7.0-103.0] vs. 18.0 [IQR 8.0-66.0] min, p = 0.833) and higher TTA (4.3 [IQR 3.7-5.1] vs. 2.9 [2.5-4.1] hours, p < 0.001). There were fewer cumulative access-site complications (1.2% vs. 9.3%, p = 0.017) and device failures (0.0% vs. 4.7%, p = 0.023), and similar same-day discharges (84.9% vs. 87.2%, p = 0.66). CONCLUSIONS: The two devices have similar hemostasis outcomes for LBVA following LAAO, but each has unique considerations. LockeT was associated with fewer access-site complications and device failures; Perclose was associated with faster TTA. Nevertheless, additional studies are still needed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Vascular closure with a suture retention device versus suture-mediated closure system after large bore venous access for electrophysiologic procedures — Science Skim