B-383 Development and Validation of a High-throughput LC-MS/MS Method for Routine Monitoring of Benzodiazepines in Serum
Michael Smith, Thomas Gorewich, Taylor Hunt, Sarah Bartock
Key points
From the paper's conclusion- Conclusion A high-throughput LC-MS/MS method was developed and validated for 5 analytes in serum using automated specimen preparation.
Abstract
Background Benzodiazepines are medications prescribed to target the central nervous system and enhance the neurotransmitter gamma-aminobutyric acid (GABA). Alprazolam, clonazepam, and lorazepam are used for short term treatment of severe anxiety, insomnia, panic disorders, and seizures. Clobazam and its metabolite norclobazam are used to treat childhood seizure disorders. Due to their risk of abuse and severe side effects, benzodiazepine concentrations are routinely monitored when prescribed. LC-MS/MS is the preferred testing method for therapeutic drug monitoring of benzodiazepines in serum due to its analytical sensitivity and specificity. For clinical laboratories, high-throughput specimen processing and analysis are also essential. Our objective was to develop and validate a high-throughput LC-MS/MS method using automated crash-and-shoot extraction of specimens after room-temperature enzymatic hydrolysis. Methods De-identified pooled serum was fortified with the 5 analytes of interest, alprazolam, lorazepam, clobazam, and norclobazam provided as reference standards from commercial sources. Automated sample preparation utilized the Hamilton Microlab STAR (Reno, NV). After serum specimens were hydrolyzed at room temperature using B-One enzyme (Kura Biotech, Los Angeles, CA) and fortified with an acetonitrile solution containing 4 internal standards (IS), they were diluted with a mix of mobile phase A (0.1% formic acid in water) and acetonitrile. Analytes were chromatically resolved using a 2.5-minute gradient on a Kinetex 2.6 µm C18 100 Å 50×3 mm column (Phenomenex, Torrance, CA). A SCIEX 4500 MS (Framingham, MA) paired with a Shimadzu Nexera LC-20AD (Columbia, MD) was used for analysis. High throughput was achieved by multiplexing 2 LC columns in staggered parallel configuration. All analytes were monitored using 2 transitions and 1 ion ratio for acceptance. Analytes were coupled with a matched IS, except for clonazepam, which used lorazepam-13C2-d4. Automated data processing was performed using Indigo Ascent software (Indigo Bioautomation, Carmel, IN). Validation studies were performed per ANSI/ASB standards and CLIA guidelines. Results are reported quantitatively for each analyte of interest. Results A quadratic 7-point calibration model was used to establish analytical measurement ranges (AMRs) of 2 to 2,000 ng/mL for alprazolam, and lorazepam and 10 to 10,000 ng/mL for clobazam and norclobazam. Within run precision ranged from 1.5% to 2.8% (n=5), and between run precision from 1.8% to 4.4% (n=25). No matrix effect was observed for dilutions at 2-, 4-, or 5-times the upper AMR limit. No carryover was observed at the upper AMR limit. Hydrolysis efficiency was 98% based B-one enzyme hydrolysis of lorazepam glucuronide at a free target concentration of 1,550 ng/mL under the same conditions used for sample preparation. Within panel and related interferences were tested at commonly observed concentrations. Except for a 51% decrease in clonazepam concentrations with nordiazepam at 5,000 ng/mL, which will be monitored during data review as a potential source of interference, no other clinically significant interferences were observed. Conclusion A high-throughput LC-MS/MS method was developed and validated for 5 analytes in serum using automated specimen preparation. This quick, simple, and effective method allows clinicians to reliably monitor these benzodiazepines in serum. The assay provides a definitive quantitative assessment for these drugs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-396 Development and validation of an efficient clinical assay for quantification of benzodiazepines in serum by LC-MS/MSClinical Chemistry · 2026
- Standardization of a Protocol to Quantify Benzodiazepines in Human Urine Samples Employing Gas Chromatography-Mass SpectrometryA Green Analytical Chemistry Approach.ACS omega · 2026
- B-381 Comparison of Benzodiazepines Positivity Rates in Serum and Urine Specimens by LC-MS/MS in a Clinical Laboratory SettingClinical Chemistry · 2026
- 1 Expanding Benzodiazepine LC-MS/MS Confirmation Improves Diagnostic Accuracy and Clinical Interpretation in Toxicology TestingAmerican Journal of Clinical Pathology · 2026
- A Validated Multi-Analyte Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Therapeutic Drug Monitoring and Targeted Toxicological Assessment of 54 Drugs and Metabolites in Human Serum.Journal of separation science · 2026
- A Multiplex LC-MS/MS Approach for Quantifying Antiepileptic Drugs for Therapeutic Drug Monitoring: Validation of 15 Representative Tests.Journal of mass spectrometry and advances in the clinical lab · 2026