A Multiplex LC-MS/MS Approach for Quantifying Antiepileptic Drugs for Therapeutic Drug Monitoring: Validation of 15 Representative Tests.
Xiaoying Tang, Ruhan Wei, Emily Chegwidden, Richard Giles, Adam M Kutnick, Drew Payto, Jessica M Colón-Franco
Department of Laboratory Medicine, Cleveland Clinic Foundation, Cleveland, OH, United States.
Key points
From the paper's conclusion- This LC-MS/MS method supports routine quantification of 15 antiepileptic drugs in a clinical laboratory using a common sample preparation and chromatography workflow across all measurements. It was operationalized into three instrument methods to accommodate the different injection volume and calibrator requirements.
Abstract
INTRODUCTION: Therapeutic drug monitoring of antiepileptic drugs is critical for optimizing clinical outcomes, minimizing toxicity, assessing drug compliance, and managing overdoses and drug interactions. OBJECTIVES: We developed and validated an LC-MS/MS method for the quantification of 15 antiepileptic drugs (ethosuximide, primidone, pentobarbital, carbamazepine-10,11 epoxide, pregabalin, gabapentin, zonisamide, lacosamide, rufinamide, felbamate, lamotrigine, topiramate, 10,11-dihydro-10-hydroxycarbamazepine, perampanel, and brivaracetam). METHODS: Antiepileptic drugs were extracted from plasma and serum using methanol-based protein precipitation, followed by dilution. A commercial ClinCal® 3-point calibrator was used for all analytes except pentobarbital, which used a 6-point in-house calibrator. Chromatographic separation was achieved using a reverse-phase C18 column with a 7.31 min elution gradient of water and methanol, both containing 2 mM of ammonium acetate. RESULTS: All analytes required a 5 µL injection volume, except ethosuximide and pentobarbital, which required 20 µL and 10 µL, respectively, for optimal performance. The method demonstrated excellent linearity for all 15 antiepileptic drugs within their respective concentration ranges, with acceptable selectivity, accuracy (80.0-118.3%), and intraday and interassay precision (CV < 7.4%). Stability studies showed that antiepileptic drugs were stable in serum and plasma for up to 24 h at room temperature, 7 days at 4 °C, and 30 days at -20 °C, and after extraction for 7 days at 4 °C. CONCLUSION: This LC-MS/MS method supports routine quantification of 15 antiepileptic drugs in a clinical laboratory using a common sample preparation and chromatography workflow across all measurements. It was operationalized into three instrument methods to accommodate the different injection volume and calibrator requirements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Validated Multi-Analyte Liquid Chromatography-Tandem Mass Spectrometry Method for Simultaneous Therapeutic Drug Monitoring and Targeted Toxicological Assessment of 54 Drugs and Metabolites in Human Serum.Journal of separation science · 2026
- Volumetric Absorptive Microsampling (VAMS) for Therapeutic Drug Monitoring of Antiseizure Medications (ASMs) in Pediatric Patients.Pharmaceuticals (Basel, Switzerland) · 2026
- Evaluation of Volumetric Absorptive Microsampling as an Alternative Matrix for Therapeutic Drug Monitoring of Antiseizure Medications: A Robust Clinical Validation Study.Therapeutic drug monitoring · 2026
- Development and validation of a LC-MS/MS method for the quantification of 22 antiepileptic drugs in saliva.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- Real-world evaluation of capillary microsampling for drug monitoring of anti-seizure medications.Epilepsia · 2026
- B-383 Development and Validation of a High-throughput LC-MS/MS Method for Routine Monitoring of Benzodiazepines in SerumClinical Chemistry · 2026