B-375 Immunoassay for the Detection of Oxycodone in Urine on Beckman Coulter AU480, AU680 and AU5800 Clinical Analyzers
Thet Wai, Ali Pilehvar, Chris Wang, Aaron Vollrath, Bernice Lin, Marie Lin
Key points
From the paper's conclusion- Conclusion The LZI Oxycodone III Enzyme Immunoassay produces accurate and reliable results across the Beckman Coulter AU480, AU680, and AU5800 clinical analyzer platforms.
Abstract
Background Oxycodone is a semi-synthetic narcotic analgesic that acts as an opioid agonist, prescribed for pain management. However, it can produce drug tolerance and therefore has the potential to be abused. This study presents the LZI Oxycodone III enzyme immunoassay for qualitative and semi-quantitative determination of oxycodone in human urine at 100 ng/mL and 300 ng/mL cutoffs. Methods The LZI Oxycodone III assay received FDA 510(k) clearance (k202007) on the Beckman Coulter AU480 clinical analyzer and was additionally validated on the AU680 and AU5800 platforms. The 510(k) studies evaluated precision, cross-reactivity, limit of detection, linearity, accuracy, and stability. All experiments were performed in qualitative and semi-quantitative modes for each cutoff. The precision study assessed assay accuracy with various drug concentrations spiked in negative human urine. Precision studies on the AU480 used samples spiked at 0 ng/mL, ±25%, ±50%, ±75%, and ±100% of the cutoff concentration. These samples were tested qualitatively and semi-quantitatively over 22 nonconsecutive days. The cross-reactivity study evaluated the interference of the assay with oxycodone, metabolites of oxycodone (oxymorphone, noroxycodone, and noroxymorphone), structurally related, and structurally unrelated compounds. Accuracy with unaltered clinical urine specimens was evaluated by method comparison studies aimed to establish the assay's reliability against established methods (LC/MS). For the method comparison study, 82 and 90 unaltered clinical samples were assessed at the 100 ng/mL and 300 ng/mL cutoffs, respectively. The LZI and Thermo Scientific DRI Oxycodone assays were also tested concurrently against a set of 79 samples for comparison. Results The assay demonstrated robust precision across all clinical analyzers. Within-run precision on the AU480 for both cutoffs yielded standard deviation (SD) values ranging from 1.1 mAU to 12.5 mAU, with a percent coefficient of variation (% CV) between 0.4% to 2.9%; total precision SDs ranged from 1.3 mAU to 20.8 mAU, with % CV from 0.8% to 5.6%. Cross-reactivity testing on the AU480 showed 100% cross-reactivity to both oxycodone and its major metabolite, oxymorphone, with consistent patterns across all analyzers. Reactivity of 43.48% and 42.86% was detected for oxymorphone-3ß-D-glucuronide at the 100 ng/mL and 300 ng/mL cutoffs. No significant interference was detected from other tested compounds. The method comparison on the AU480 achieved over 90% positive and negative agreement across both the 100 ng/mL and 300 ng/mL cutoff with similar results across all AU platforms. Discrepant samples were only found near the cutoff range (±50% cutoff concentration) across all AU platforms. In addition, the LZI Oxycodone III Assay was found to have fewer total discrepant samples as compared to the Thermo Scientific DRI Oxycodone assay when compared against LC/MS. Conclusion The LZI Oxycodone III Enzyme Immunoassay produces accurate and reliable results across the Beckman Coulter AU480, AU680, and AU5800 clinical analyzer platforms. The assay is sensitive in detecting oxycodone with low interference from other opiates.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-376 Immunoassay for the Detection of Fentanyl in Urine on Beckman Coulter AU5800, AU680, AU480, DxC700 AU, and DxC500 AU Clinical AnalyzersClinical Chemistry · 2026
- B-374 Immunoassay for the Detection of Xylazine in Urine on Beckman Coulter AU480 AnalyzerClinical Chemistry · 2026
- B-401 Evaluating the Value of a Dedicated Hydrocodone Immunoassay Screen in Routine Urine Drug TestingClinical Chemistry · 2026
- Comparison of Immunoassay Results near Cutoff Values with LC-MS/MS in Individuals with Substance Use Disorder Under Probation.Metabolites · 2026
- A-379 Comparative Performance Evaluation of Two Multi-Drug Urine Screening Cups with Fentanyl DetectionClinical Chemistry · 2026
- 49 Assay Performance and Test Utilization: A Comparison of E-Z Cup Point-of-care Device and Alinity c Automated Urine Drug Screening PlatformsAmerican Journal of Clinical Pathology · 2026