American Journal of Clinical Pathology

70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting Issues

Imad Tarhoni, Kelsey Reschly, Mackenzie Edward, Margaret Gearhart, Nichole Korpi-Steiner, Steven Cotten

Key points

From the paper's conclusion
  • This study evaluated the D-100’s performance in a cohort enriched for hemoglobinopathies, focusing on its ability to report accurate HbA1c results with potential hemoglobin variants.

Abstract

Background The Bio-Rad D-100 is a cation-exchange HPLC-based HbA1c instrument designed for routine high throughput analysis. However, cation-exchange HPLC methods may be unpredictable in the presence of hemoglobin variants. This study evaluated the D-100’s performance in a cohort enriched for hemoglobinopathies, focusing on its ability to report accurate HbA1c results with potential hemoglobin variants. Methods Accuracy was assessed using 10 NGSP-certified samples, and flag rates were evaluated in two cohorts: a routine cohort (n = 221) and an enriched hemoglobin variant cohort (n = 34). The output parameters (A0%, F%, and S-window%) were compared with the same parameters on the Bio-Rad VARIANT™ II Hemoglobin Testing System. These parameters were categorized as “Low total area/ No HbA1c peak/ No Result,” “F > 5%, Peak in S-Window,” or “Peak in S-Window,” and contrasted with results from the VARIANT II for S-dependent variants (HbSS, HbSC, HbS-β⁺ thalassemia, HbS-β0 thalassemia). Results NGSP specimens strong agreement with assigned values (Deming slope = 1.064, intercept = −0.37, R = 0.9997), with a mean bias of 1.93%. In the routine cohort (n = 221), 5% of samples were flagged as “total area too low, should not report HbA1c,” primarily at low HbA1c concentrations (4–5%). Interestingly, 2 of 10 NGSP samples were flagged similarly, yet generated accurate results. In the enriched hemoglobin variant cohort (n = 34), all samples were flagged as unsuitable for reporting, highlighting a limited ability to report results in variant-rich populations. Among S-related variant disease cases (n = 32), 12 were correctly classified as “F > 5%, Peak in S-Window,” and 11 as “Possible Variant Interference,” while 19 showed “Low total area, No HbA1c Peak/ No Result,” with no S-window or F-area peaks. The Bio-Rad VARIANT™ II identified prominent peaks in these cases. Two non-S variant samples confirmed by the VARIANT II were flagged as unsuitable for reporting by the D100 for HbA1c, but instead were classified as “S-window peaks”. Additionally, HbA1c values were generated in samples with high S and F areas, revealing a failure to differentiate HbA1c results accurately in the presence of a potential interference. Reproducibility of interference detection using multiple injections was consistent for 9 out of 11 samples, while 2 were flagged as “low total area” in the first run and showed “no HbA1c” in the second. Two additional samples were repeated up to 4 times, showing inconsistent results, with one flagged for “possible variant.” Conclusion The Bio-Rad D-100 demonstrates predictable analytical performance with NGSP samples and variant-free specimens. However, variability exists for detection at low HbA1c concentrations and generates nonspecific flagging in hemoglobin variant samples, particularly HbSS. Laboratories should evaluate HbA1c interference flags and implement confirmatory strategies when using cation-exchange HPLC methods in variant-rich populations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting Issues — Science Skim