B-336 ICP MS analysis of cellulose based microsamples containing 10 or 50 µL of blood

Donald Chace, Charles Harper, Jerry Vockley, Nesta Bortey-Sam

Key points

From the paper's conclusion
  • We have found that the baseline lead in cellulose (a feature of all cellulose based dried microsamples) is low for all devices when only low lead lots of paper are chosen.

Abstract

Background We completed a large study in 2025 with the University of Pittsburgh where we analyzed Pb in the blood of newborns and others using a 50 µL Capitainer B50 microsampling device. We have also validated a 10 µL Captitainer B10 device. The Capitainer devices were chosen because they have a unique configuration where the cellulose disk is protected from contamination. It is well known that Pb is in cellulose paper and must be accounted for. We have developed a robust method to achieve excellent precision of measurement and control for background making it a more suitable device for Pb monitoring than the current filter paper applications.. Methods Dried blood collected on cellulose disks using Captitainer devices are extracted with 3 mL of UHP water containing 0.01% Triton X-100. They are vortexed then sonicated, 30 min, platform shaken, 30 min. After these steps, 3 mL of 1% nitric acid is added to produce a final volume of 6 mL volume. The tube vortexed then left standing, 30 min. A Thermo Fisher iCAP RQ plus ICP MS with an Elemental Scientific 4DX autosampler / FAST system was used to analyze 2 mL of extract in approximately 1 min. Results We developed our assay from water, plasma, and whole blood to optimize our sensitivity, accuracy and precision. We used Bio Rad QC blood at 3 Pb enriched levels and participated in 2 proficiency testing programs. The results show excellent correlatio between liquid and dried with an r-squared of 0.99 and a slope of near one for all comparson studies. We have found that the baseline lead in cellulose (a feature of all cellulose based dried microsamples) is low for all devices when only low lead lots of paper are chosen. With this control and rigorous manufacturing of the cards, the background levels are consistent and reproducible. The limits of detection for both devices is much less than 1 µg/dL. Samples were diluted 600 fold for B10 and 120 fold for B50, therefore the best matrix match is water. Three levels of Bio Rad blood standards for lead was used as a routine control and to insure that the standard curve was accurate. Regarding the study of newborns and moms (∼1800 samples) we found that newborns had a lower median Pb value of 0.57 compared to adult mothers which was 0.75. The Fe concentrations of newborns were nearly double (1.8x) than their adult moms, reflecting the higher hematocrit of newborns. In the Pitt study it was clear that Capitainer cards (B50) did not interfere with the collection of newborn screening blood cards. Logistically, it was determined that cards are more easily handled post collection as they do not require drying in an open-air environment risking exposure to contamination. These studies show that Capitainer devices can be used as an alternative device for "Guthrie cards" and also an alternative to venous blood for screening purposes, especially in Pediatric applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers