B-334 A preliminary assessment of a recombinant cardiac troponin for use in understanding the differences in the quantities reported by in-vitro diagnostic test devices
Miriam Bialojahn, Claudia Swart, Cristian-Gabriel Arsene, Gavin O’Connor
Key points
From the paper's conclusion- Preliminary tests have shown observable differences between phosphorylated and unphosphorylated cTnI, suggesting the possible impact of protein phosphorylation on the antibody based in-vitro diagnostic measurement devices. [1] Reference Institute for Bioanalytics (2023).
Abstract
Background Cardiac troponin I (cTnI) is used as a biomarker for myocardial infarction and other heart diseases. It is routinely measured using immunoassays based on different antibodies. However, it has been reported that different diagnostic platforms give different quantitative results for the same samples[1]. These differences can stem from the different epitopes recognized by the different antibodies used by manufactures of in-vitro diagnostic tests. An alternative cause of these discrepancies are the complexation of cTnI with the different forms of troponin present in the samples, or post translational modifications (PTMs). One such variation is the phosphorylation of the amino acids in the sequence. These can impact the binding efficiency of antibodies if they are located inside the epitope, by steric hindrance or by changing the 3-dimensional conformation of the protein. Methods A phosphorylated recombinant cTnI material was produced by Goethe University, Frankfurt, using protein kinase A. The degree of phosphorylation was determined by measuring the mass of the intact protein using an electrospray ionization source under native spray conditions with an Orbitrap MS. To determine the individual sites of phosphorylation, a tryptic digest was measured using capillary liquid chromatography coupled with a trapped ion mobility time of flight (timsTOF) mass analyzer with an electrospray ion source. The data were analyzed using Peaks® software to identify the most significant sites of phosphorylation. Preliminary tests to investigate the impact of phosphorylation on the binding efficiency of antibodies were performed using isotope dilution high performance liquid chromatography coupled with inductively coupled plasma mass spectrometry (HPLC-ICP-MS), to compare the amount of substance ratio of unphosphorylated recombinant cTnI with phosphorylated cTnI. Anti-cTnI antibody RecChim19C7 (HyTest Ltd.) was coupled to protein G magnetic microbeads to act as capture antibody. For detection, anti-cTnI antibody 84 (HyTest Ltd.) was labeled using the MaxPar® X8 antibody labeling kit (Standard BioTools Inc.) with 162Dy for unphosphorylated and 163Dy for phosphorylated cTnI. Results The results obtained on the Orbitrap MS suggested the presence of at least 4 different variants in the phosphorylated recombinant material. The observed spectrum suggested a amount-of-substance ratio of 4:3:2:1 for the site occupancy of 3, 4, 2 and 5 phospho groups. The analysis performed with Peaks® software suggested the presence of 9 highly significant phosphorylation sites, including 8 previously found sites (S23, S24, T51, T129, T143, S150, S166, S199) and one novel site (T123). The preliminary tests using isotope dilution HPLC-ICP-MS showed a difference of about 20% between the phosphorylated and unphosphorylated cTnI. Conclusion The phosphorylated recombinant material has 9 phosphorylation sites, of which several lie either directly in or close to the epitopes of commonly used antibodies for the routine detection of cTnI. This makes it suitable for future use in the investigation of the impact on the binding efficiency of different antibodies. Preliminary tests have shown observable differences between phosphorylated and unphosphorylated cTnI, suggesting the possible impact of protein phosphorylation on the antibody based in-vitro diagnostic measurement devices. [1] Reference Institute for Bioanalytics (2023). Results of an interlaboratory comparison (PT scheme 3/23) for cTnI. https://www.rfb.bio/cgi/displayAnaStats?rv_type=all&rvTypeForDetails=CM&year=2023&rv_nu
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Single-step determination of cardiac troponin molecular form ratios using multiparametric SH-SAW biosensing.Talanta · 2026
- A-023 Evaluation of Interfering Substances in VITROS High-Sensitivity Cardiac Troponin I AssayClinical Chemistry · 2026
- A-375 Verifying analytical performance of i-STAT high sensitivity troponin cartridges in an urban academic medical center emergency departmentClinical Chemistry · 2026
- A-046 A novel cardiac troponin assay circumvents positive and negative macrotroponin interferenceClinical Chemistry · 2026
- Long forms of cardiac troponin T for myocardial infarction diagnosis: the SuperTROPO studyEuropean Heart Journal · 2025
- A-002 Evaluation of the Abbott i-STAT devices for both high-sensitivity cardiac troponin I and creatinineClinical Chemistry · 2026