A-389 Understanding sample-to-sample variability in capillary blood collections: comparison with venous sampling
Kaylan Olds, J Poppa, Spencer Stowell, Evgenia Macke, Brian Massie, Vanaja Bodeddula, Sri Madireddy, Erwin Berthier,
Key points
From the paper's conclusion- Against this benchmark, capillary collection by HCPs demonstrated lower variability than venous sampling for lipid analytes CHOL, and HDL, with SD ratios of 0.44-0.73 (confidence intervals entirely below 1.0).
Abstract
Background Capillary blood collection devices offer a less invasive alternative to venipuncture, however, sample-to-sample variability within and between operators remains a critical performance metric. Unlike venous draws where large volumes enable multiple operators to analyze aliquots from the same sample—which allow evaluation of isolated operator effects—capillary volumes require separate collections per operator, confounding operator technique with inherent physiological sample-to-sample variation. To account for this limitation, a study was designed to compare sample-to-sample variability in capillary collections to that observed using two consecutively collected venous specimens drawn during the same venipuncture procedure, thus establishing a baseline for sample-to-sample variability against which capillary operator variability could be contextualized. This study evaluated inter-operator variability among HCPs and intra-operator variability among lay user self-collections using the Tasso+ Serum Gel Capillary Blood Collection System compared with venous sample-to-sample variability. Methods For inter-operator assessment, 80 participants each had capillary samples collected by two different trained HCPs using the Tasso+ device. For intra-operator assessment, 45 lay users performed duplicate self-collections. Two separate venous samples drawn during a single venipuncture procedure served as the reference benchmark for both analyses. Seven analytes were evaluated: total cholesterol (CHOL), creatinine (CREAT), HDL, high-sensitivity C-reactive protein (hsCRP), LDL, lipoprotein(a) [LP(a)], and triglycerides (TRIG). Within-subject standard deviations (SD), coefficients of variation (CV), and SD ratios (capillary/venous) with 95% confidence intervals were calculated. Results HCP Inter-operator variability for capillary samples demonstrated excellent reproducibility with mean CVs of 0.5% (CHOL, LDL), 0.8% (HDL), 1.5% (CREAT), 1.6% (hsCRP), 1.9% (TRIG), and 2.5% [LP(a)]. Capillary-to-venous (reference) SD ratios (95% CIs) demonstrated very high reproducibility between capillary samples for lipid markers: CHOL 0.73 (0.58-0.89), LDL 0.55 (0.41-0.69), and HDL 0.44 (0.31-0.57). CREAT showed a ratio of 0.87 (0.57-1.17), whereas hsCRP 1.37 (0.91-1.84), LP(a) 1.32 (0.57-2.07), and TRIG 1.75 (1.37-2.13) demonstrated ratios above 1.3. Lay user intra-operator variability for self-collected capillary samples showed mean CVs of 0.7% (CHOL, 1.0% (HDL), 1.7% (CREAT), 2.3% (TRIG), and 3.0% [LP(a)]. Self-collected capillary-to-venous (reference) SD ratios were: CHOL 0.91 (0.66-1.17), LDL 0.83 (0.44-1.21), HDL 0.54 (0.34-0.74), CREAT 1.05 (0.70-1.39), hsCRP 1.02 (0.71-1.34), LP(a) 1.31 (0.24-2.38), and TRIG 2.22 (1.52-2.92). Conclusion Capillary sample-to-sample variability was similar to, or lower than, observed for venous draw-to-draw replicates for nearly all analytes assessed. Importantly, this study identified that even venous "gold standard" methods exhibit measurable sample-to-sample variation, indicating that measurable between-sample variation exists in both venous and capillary sampling. Against this benchmark, capillary collection by HCPs demonstrated lower variability than venous sampling for lipid analytes CHOL, and HDL, with SD ratios of 0.44-0.73 (confidence intervals entirely below 1.0). For hsCRP and LP(a) the SD ratios were above 1.0 with CIs spanning across 1.0, and for TRIG, the ratios and CIs were greater than 1.30. Self-collection showed similar patterns and largely remained within clinically acceptable limits. This is a significant finding that challenges conventional assumptions around reproducibility of venous drawn samples and suggests this source of variation should be taken into consideration when evaluating sample-to-sample capillary collections for reproducibility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Capillary Blood Collection Methods: An assessment of different collections technologies compared to traditional venous sampling.Annals of clinical biochemistry · 2026
- A-390 Method comparison of a novel capillary blood collection device for clinical chemistry testing: comparison with venous blood on two automated platformsClinical Chemistry · 2026
- B-021 Evaluation of Capillary Serum as an Alternative Matrix for Hepatic Function TestsClinical Chemistry · 2026
- A-391 Short-term stability of clinical chemistry analytes in serum samples collected using a novel capillary blood collection deviceClinical Chemistry · 2026
- Automatic for the people: an introduction to self-collected capillary blood testingJournal of Patient Centricity · 2026
- Capillary blood in core laboratories: current and future challengesClinical Chemistry and Laboratory Medicine (CCLM) · 2025