B-275 Operational and Environmental Benefits of Load-and-Go Liquid QC in Clinical Chemistry
Yuna Choi, Sollip Kim, Woochang Lee, Sail Chun
Key points
From the paper's conclusion- Based on Asan Medical Center consumption data, InteliQ corresponded to an approximate 40% reduction in chemistry consumables and 60% reduction in immunoassay and cardiac consumables.
Abstract
Background Internal quality control (QC) is essential for ensuring reliability of clinical laboratory testing. Conventional lyophilized QC materials require reconstitution, aliquoting, freezing, and thawing, increasing manual workload, error risk, and consumable waste. Tube-based liquid QC systems with integrated barcoding (InteliQ Assayed Multiqual Control, Bio-Rad Laboratories) have been introduced to streamline workflows. However, real-world stability data and operational and environmental impacts remain limited. We evaluated open-vial stability and workflow efficiency of InteliQ and performed scenario-based simulations of lot crossover costs and laboratory waste versus conventional practice. Methods Four components were assessed: stability, workflow efficiency, crossover-cost, and waste estimation. The evaluation included 29 general chemistry analytes measured on Roche Cobas 8000 and Beckman Coulter AU5822, CRP and hsCRP measured on both Cobas 8000 and AU5822, and four cardiac markers (BNP, CK-MB, myoglobin, and troponin I) measured on the Siemens Atellica IM. For stability testing, InteliQ Assayed Multiqual, Immunology, and Cardiac controls were handled according to the manufacturer’s instructions. Unopened vials were stored frozen, thawed at room temperature before first use, and recapped and stored at 2–8°C between daily runs. Analytes were measured in triplicate from Day 0 to the manufacturer-stated open-vial stability limits (2 days for CO2; 7 days for others). Percentage variation from baseline was evaluated against analytical performance specifications defined by the Royal College of Pathologists of Australasia. Workflow efficiency was evaluated by timing each procedural step for conventional versus tube-based QCs, comparing execution time and handling steps. For crossover-cost simulation, annual costs were estimated according to QC lot change frequencies of 2, 3, and 4 changes. Because reagent pricing and labor costs vary across institutions, a standardized cost model based on current Korean health insurance reimbursement rates was applied. For waste estimation, we assumed one analyzer operating 250 days per year under 2–4 daily QC run scenarios. Consumables were weighed/counted, and annual waste was projected using reduction rates from Asan Medical Center QC consumption data. Results On the AU5822, open-vial stability met specifications for all analytes except total CO2, which failed on Day 2. On the Cobas 8000, most analytes met 7-day claims, with exceptions including phosphorus (Day 7), sodium (Day 5), and total CO2 (Day 2). Conventional QC workflow required 3,165 seconds versus 1,691 seconds using InteliQ, a 46.6% reduction, primarily due to elimination of reconstitution and aliquoting and simplified handling. Crossover-cost simulation estimated annual costs according to lot change frequency: approximately USD 4,680 for two, USD 9,360 for three, and USD 14,040 for four. Based on Asan Medical Center consumption data, InteliQ corresponded to an approximate 40% reduction in chemistry consumables and 60% reduction in immunoassay and cardiac consumables. Applied to annual operation scenarios, projected total waste reductions were 81.1% (2 runs/day), 81.7% (3 runs/day), and 80.9% (4 runs/day). Conclusion InteliQ demonstrated satisfactory open-vial stability for most analytes, although total CO2 showed limited stability, requiring careful handling. Beyond workflow improvements, tube-based QC systems may help mitigate costs associated with frequent lot crossovers and substantially decrease laboratory waste.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation and application of automated quality control in clinical coagulation testing.Research and practice in thrombosis and haemostasis · 2026
- Risk-based redesign of internal quality control procedures for emergency immunoassays in clinical laboratories: a multicenter study from China.Frontiers in medicine · 2026
- Internal quality control for immunodiagnostic laboratories.Asian journal of transfusion science · 2026
- A-307 Determining risk-based internal quality control frequency for biochemical testing using Sigma metrics and the QC Constellation toolClinical Chemistry · 2026
- Application of Six Sigma, Quality Goal Index, Method Decision Chart, and Operating Specification Metrics as Stringent Quality Control Tools for Assessing the Analytical Performance of BUN, Creatinine, and Glucose in an ISO 15189:2022–Accredited LaboratoryThe Journal of Applied Laboratory Medicine · 2026
- Integration of Patient-Based Real-Time Quality Control with Conventional Internal Quality Control: Improved Error Detection and Cost-Benefit Analysis in a Clinical Laboratory.EJIFCC · 2026