Application of Six Sigma, Quality Goal Index, Method Decision Chart, and Operating Specification Metrics as Stringent Quality Control Tools for Assessing the Analytical Performance of BUN, Creatinine, and Glucose in an ISO 15189:2022–Accredited Laboratory
Maisoon Amarni, Kamal Dumaidi, Sara Abbadi, Nataly Fakhouri, Amer Al-Jawabreh
Arab American University
Key points
From the paper's conclusion- This study evaluated the analytical performance of 3 routine analytes, blood-urea nitrogen (BUN), creatinine, and glucose using Sigma metrics, visualized through the Method Decision Chart (MDC), Quality Goal Index (QGI), and Operating Specifications (OPSpecs) charts, to recommend appropriate internal quality control (IQC) procedures at 3 levels (L1, L2, L3).
Abstract
BACKGROUND: Reliable analytical performance is vital for accurate diagnosis in clinical chemistry laboratories. Quality control (QC) strategies must be customized to assess performance, meet regulatory standards, and optimize resource use. This study evaluated the analytical performance of 3 routine analytes, blood-urea nitrogen (BUN), creatinine, and glucose using Sigma metrics, visualized through the Method Decision Chart (MDC), Quality Goal Index (QGI), and Operating Specifications (OPSpecs) charts, to recommend appropriate internal quality control (IQC) procedures at 3 levels (L1, L2, L3). METHODS: A 6-month retrospective analysis (June-November 2024) was performed in an ISO 15189:2022 accredited primary health care laboratory in Jenin District, Palestine. Internal QC and External Quality Assessment Scheme data were used to calculate the coefficient of variation (%CV), %bias, and Sigma values across all IQC levels, applying CLIA total allowable error limits. For analytes with Sigma <4, QGI ratios differentiated between imprecision, inaccuracy, or combined error sources. MDC and OPSpecs charts guided the selection of Westgard QC rules. RESULTS: Glucose demonstrated excellent performance (σ > 6), allowing a simplified IQC protocol (1-3.5s, N3, R1). Creatinine showed moderate performance (σ = 3.1-13.6) requiring multirule application (2 of 3-2s, R4s) at Level 2. BUN displayed poor Sigma values (σ < 4), with QGI analysis indicating alternating imprecision and inaccuracy; OPSpecs recommended full multirule implementation and root-cause investigation. CONCLUSION: The combined use of Sigma metrics, QGI, MDC, and OPSpecs charts offers an effective, risk-based QC framework enabling early detection, continuous monitoring, and timely correction of analytical errors in ISO 15189-certified laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-307 Determining risk-based internal quality control frequency for biochemical testing using Sigma metrics and the QC Constellation toolClinical Chemistry · 2026
- Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory.Laboratory medicine · 2026
- Longitudinal Sigma-Metric Profiling of Two Clinical Chemistry Platforms Using Peer-Group Internal Quality Control.Diagnostics (Basel, Switzerland) · 2026
- B-266 A multisite review across the USA of major competitive biochemistry platforms reveals significant differences in analytical performanceClinical Chemistry · 2026
- B-265 A Large Multisite Study of Beckman Coulter clinical chemistry performance demonstrates a significant majority of assays are eligible to use less QCClinical Chemistry · 2026
- Improving Internal Quality Control Through Metrological Reinterpretation of Westgard RulesMetrology · 2026