B-195 Preliminary implementation of mass spectrometry (EXENT®) for the detection of oligoclonal bands in CSF
Nick Paciaroni, Marta Outón, José Luis Veiga, Mónica Pascual, Raúl Vidal, Teresa Cedena, Joaquín Martínez, Luisa María Villar,
Key points
From the paper's conclusion- Conclusion In this preliminary evaluation, EXENT® mass spectrometry showed excellent agreement with IEF for IgG OCB-positive and negative classifications.
Abstract
Background Detection of IgG oligoclonal bands (OCBs) in cerebrospinal fluid (CSF) by isoelectric focusing (IEF) with immunoblotting remains the reference laboratory method for demonstrating intrathecal immunoglobulin synthesis in multiple sclerosis (MS). Although analytically sensitive, IEF is labor-intensive, qualitative, and subject to interobserver variability. Mass spectrometry–based immunoglobulin profiling using the EXENT® Analyzer enables molecular-level detection of clonal intact light chains (iLC) through selective immunocapture and MALDI-TOF analysis, providing a digital and potentially more standardized readout. Methods Twelve paired CSF–serum samples previously characterized by IEF in a reference center were analyzed. Patterns included three mirror cases, three IgG+IgM positive cases, three IgG positive cases, and three negative cases. CSF samples were processed without external dilution; serum was analyzed both diluted and undiluted, with improved analytical performance observed in undiluted serum. The EXENT® workflow applies an internal ∼1:10 dilution during analytical processing. Independent expert interpretation was performed for both methodologies. Results EXENT® demonstrated 100% concordance with IEF for IgG OCB-positive and negative cases. Mirror-pattern cases showed limited agreement, with 1/3 fully concordant (33%). In one mirror case, only serum peaks were detected; in another, no differential signal was observed. Mirror-pattern cases were not associated with intrathecal synthesis or clinical MS diagnosis, consistent with their biological interpretation. These findings underscore the importance of appropriate alignment of serum and CSF immunoglobulin concentrations to minimize potential misclassification and preserve diagnostic specificity. No IgM oligoclonal patterns were detected by EXENT®, including samples classified as IgG+IgM positive by IEF. Analytical review indicated that IgG—and particularly IgM—concentrations in CSF were frequently near the system’s analytical limit of detection. The internal dilution step may reduce the effective concentration of intact light chains available for immunocapture, potentially influencing analytical sensitivity, especially for IgM detection. Conclusion In this preliminary evaluation, EXENT® mass spectrometry showed excellent agreement with IEF for IgG OCB-positive and negative classifications. However, reduced concordance in mirror patterns and absence of IgM detection suggest that analytical sensitivity in low-concentration CSF samples may be influenced by internal dilution. Further optimization of CSF-specific workflows and larger multicenter studies with formal statistical validation are needed to better define the role of mass spectrometry in CSF oligoclonal band assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-196 Comprehensive clinical validation of the Immunoglobulin Isotypes (GAM) assay for the EXENT® Analyzer assay in monoclonal gammopathiesClinical Chemistry · 2026
- B-039 Harmonizing M-protein analysis: an international, multi-center reproducibility study of the EXENT® System and evaluation of concordance to electrophoretic methodsClinical Chemistry · 2026
- Performance evaluation of an automated mass spectrometry system for the measurement and isotyping of monoclonal immunoglobulinsClinica Chimica Acta · 2026
- B-358 Revisiting Biclonal Gammopathies in the Era of High-Resolution Mass Spectrometry with EXENT®Clinical Chemistry · 2026
- Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.Clinical chemistry and laboratory medicine · 2026
- B-064 Bridging Quantitative MALDI-TOF Mass Spectrometry to IMWG Complete Response: Deriving a Serum Cut-off Equivalent to Negative ImmunofixationClinical Chemistry · 2026