B-173 Impact of Cut-off Adjustment on Inter-Assay Agreement of Anti-Cardiolipin and Anti-ß2 Glycoprotein I IgG Across Three Automated Immunoassay Platforms
Howon Lee
Key points
From the paper's conclusion- Platform-specific cut-off adjustment for aCL IgG improves specificity and inter-assay concordance while preserving acceptable sensitivity, underscoring the need for assay-specific harmonization strategies in routine clinical laboratories.
Abstract
Background Anti-cardiolipin (aCL) and anti-ß2 glycoprotein I (aß2GPI) IgG assays are essential in the evaluation of systemic lupus erythematosus (SLE) and antiphospholipid antibody testing. With increasing adoption of automated immunoassays, concerns have emerged regarding inter-assay variability and overclassification of low-titer positivity, particularly when uniform cut-off values are applied across different platforms. This study compared the diagnostic performance of three automated immunoassays and assessed the effect of cut-off modification on assay concordance. Methods Serum samples from 248 patients (81 SLE and 167 non-SLE) were analyzed for aCL IgG and aß2GPI IgG using three automated platforms: a chemiluminescent immunoassay (BIO-FLASH®), a multiplex bead-based immunoassay (BioPlex® 2200), and a fluorescence enzyme immunoassay (Phadia® 250). Sensitivity and specificity for SLE were evaluated using a uniform cut-off of 20 units. For BIO-FLASH® aCL IgG, additional analyses were performed with elevated cut-offs of 30 and 40 units to assess changes in diagnostic performance and inter-assay concordance. Results At a cut-off of 20 units, marked inter-assay differences were observed for aCL IgG. Sensitivity for SLE was highest on BIO-FLASH® at 62% (50/81), compared with 31% (25/81) on BioPlex® 2200 and 36% (29/81) on Phadia® 250. However, specificity for aCL IgG was lowest on BIO-FLASH® (69%, 115/167), whereas BioPlex® 2200 and Phadia® 250 demonstrated higher specificities of 95% (159/167) and 88% (147/167), respectively. For aß2GPI IgG at the same cut-off, sensitivity ranged from 31% to 49% (BIO-FLASH® 49%, 40/81; BioPlex® 2200 41%, 33/81; Phadia® 250 31%, 25/81), with relatively consistent specificity across platforms (87–91%). Raising the BIO-FLASH® aCL IgG cut-off resulted in a stepwise trade-off between sensitivity and specificity. Sensitivity decreased from 62% at 20 units to 43% (35/81) at 30 units and 33% (27/81) at 40 units, while specificity increased from 69% to 80% (134/167) and 87% (145/167), respectively. Elevation of the cut-off substantially reduced low-titer positivity among non-SLE patients and improved concordance with the other platforms. Conclusion Low-titer aCL IgG results are a major source of inter-assay discordance among automated immunoassays. Application of uniform cut-offs across platforms may lead to reduced specificity and potential overclassification, particularly with CLIA-based assays. Platform-specific cut-off adjustment for aCL IgG improves specificity and inter-assay concordance while preserving acceptable sensitivity, underscoring the need for assay-specific harmonization strategies in routine clinical laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic accuracy of cell-bound complement activation product assays for systemic lupus erythematosus: a systematic review and marker-stratified bivariate meta-analysis.Frontiers in immunology · 2026
- Interval-specific likelihood ratios improve interpretation of antiphospholipid antibody testing across different analytical platforms.Clinical chemistry and laboratory medicine · 2026
- B-148 A Comparative Performance Evaluation of Two Different CIC C1q ELISA AssaysClinical Chemistry · 2026
- Impact of testing method and analyte on the diagnosis of antiphospholipid syndrome: a survey by the association of medical laboratory immunologists.Clinical chemistry and laboratory medicine · 2026
- Advances and pitfalls in anti-double-stranded DNA autoantibody detection and interpretation.World journal of biological chemistry · 2026
- Analytical Comparison of an In-House FARR Fluorescence Immunoassay with Commercial Analytical Assays for the Detection of Anti-Double-Stranded DNA Antibodies.Diagnostics (Basel, Switzerland) · 2026