B-148 A Comparative Performance Evaluation of Two Different CIC C1q ELISA Assays

Jaehoon Choi, Gwiyoung Oh

Key points

From the paper's conclusion
  • Conclusion Bühlmann and DiaMetra CIC C1q ELISA assays demonstrate negligible quantitative correlation (r = 0.073) and poor qualitative agreement (kappa = 0.145), indicating substantial methodological differences despite similar ELISA-based principles.

Abstract

Background Circulating immune complexes (CIC) containing C1q are important biomarkers for monitoring disease activity in autoimmune disorders such as systemic lupus erythematosus (SLE) and rheumatoid arthritis (RA). Two commercially available ELISA assays for CIC C1q quantification, Bühlmann CIC ELISA (Bühlmann Laboratories AG, Switzerland) and DiaMetra CIC C1q ELISA (Dia.Metra Srl, Italy) are widely used in clinical laboratories. However, their analytical and clinical comparability has not been thoroughly evaluated. This study aimed to assess the correlation and agreement between these two assays. Methods A total of 127 serum samples were analyzed using both Bühlmann and DiaMetra CIC C1q ELISA kits over four independent runs (December 19, 2024; January 9, 2025; February 4, 2026; February 11, 2026). Testing was performed according to each manufacturer's instructions. Results were classified as negative, equivocal, or positive based on manufacturer recommended cut-offs: Bühlmann (<3.2, 3.2-5.0, >5.0 µg Eq/mL) and DiaMetra (<16, 16-18, >18 µgEq/mL). Statistical analyses included Pearson and Spearman correlation coefficients, linear regression, and Cohen's kappa for categorical agreement. Results Of 127 samples, Bühlmann produced 29 (22.8%) below-limit-of-detection (<1.00 µg Eq/mL) and 1 (0.8%) above-upper-limit (>50 µg Eq/mL) results, with 97 (76.4%) quantifiable values (mean ± SD: 5.55 ± 6.47 µg Eq/mL; range: 1.00-43.86). DiaMetra yielded 125 quantifiable results (mean ± SD: 8.63 ± 10.14 µgEq/mL; range: 1.62-61.30). Correlation analysis on 96 paired quantifiable results revealed extremely poor agreement:Pearson correlation: r = 0.073 (p = 0.481)Spearman correlation: r = 0.091 (p = 0.376)Linear regression: DiaMetra = 0.128 × Bühlmann + 8.362 (R² = 0.0053) Qualitative classification showed marked discordance. Bühlmann classified 66 (52.0%), 31 (24.4%), and 30 (23.6%) samples as negative, and positive, respectively. DiaMetra classified 110 (86.6%), 2 (1.6%), and 13 (10.2%) samples as negative, respectively. Among 93 samples excluding equivocal results, overall categorical agreement was 71.0% (66/93) with Cohen's kappa of 0.145 (slight agreement). Using DiaMetra as reference, Bühlmann demonstrated sensitivity of 55.6% (5/9) and specificity of 72.6% (61/84). Notably, 23 of 84 (27.4%) DiaMetra-negative samples were classified as positive by Bühlmann. Conclusion Bühlmann and DiaMetra CIC C1q ELISA assays demonstrate negligible quantitative correlation (r = 0.073) and poor qualitative agreement (kappa = 0.145), indicating substantial methodological differences despite similar ELISA-based principles. The 5.6-fold difference in manufacturer cut-offs (Bühlmann: 3.2 vs. DiaMetra: 18 µgEq/mL) and high false-positive rate of Bühlmann relative to DiaMetra suggest these assays are not interchangeable. Laboratories must establish method-specific reference intervals and cannot directly compare results between methods. The lack of correlation raises concerns about CIC C1q assay standardization and highlights the urgent need for internationally certified reference materials and harmonized calibration procedures in autoimmune disease monitoring.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-148 A Comparative Performance Evaluation of Two Different CIC C1q ELISA Assays — Science Skim