Clinical chemistry and laboratory medicine

Impact of testing method and analyte on the diagnosis of antiphospholipid syndrome: a survey by the association of medical laboratory immunologists.

Rohan Willis, Vijayalakshmi Nandakumar, Vijaya Murthy, Ken Dier, May Y Choi, Susan L Fink, Marvin J Fritzler, Melissa R Snyder,

Antiphospholipid Standardization Laboratory, Division of Rheumatology, University of Texas Medical Branch and Louisville APL Diagnostics, Galveston, TX, USA. · ARUP Institute for Clinical and Experimental Pathology and Department of Pathology, University of Utah School of Medicine, Salt Lake City, UT, USA.

Key points

From the paper's conclusion
  • Generally, the non-ELISAs showed better performance for selected analytes, particularly aCL IgG, while ELISAs demonstrated greater consistency. Although agreement between methods varied for specific antiphospholipid antibodies, overall diagnostic accuracy was comparable when the complete antiphospholipid antibody profile was assessed. Clinicians should consider these factors to ensure accurate and reliable diagnostics.

Abstract

OBJECTIVES: To evaluate and compare the performance, variation, and agreement of ELISA and non-ELISA methods for detecting anti-cardiolipin (aCL) and anti-beta 2 glycoprotein 1 (aβ2GP1) antibodies across clinical laboratories and assess diagnostic standardization. METHODS: ELISA and non-ELISA methods for IgG, IgM, and IgA aCL and aβ2GP1 antibodies were evaluated in patients with antiphospholipid syndrome (APS), APS-like disease, and healthy controls. Agreement (κ and Spearman correlation), sensitivity, specificity, area under the curve (AUC), and relative risk (RR) for APS, thrombotic events (TE), and pregnancy-related morbidity (PRM) were assessed. RESULTS: Correlations were generally stronger within ELISA and within non-ELISA than between ELISA and non-ELISA methods, although substantial inter-assay variability existed within both groups. For APS diagnosis, IgG aCL non-ELISAs demonstrated higher sensitivity than ELISAs (0.73-0.80 vs. 0.50-0.63; p=0.026) with similar specificity (0.96-0.99 vs. 0.94-1.00). AUCs for IgG aCL and IgG aβ2GP1 were comparable across individual assays regardless of method (0.83-0.93 and 0.76-0.94, respectively). Non-ELISA IgG aCL assays also showed higher RR estimates for APS (8.3-12.0 vs. 5.7-7.4; p=0.024). However, when aCL and aβ2GP1 results were assessed as a complete antibody profile, sensitivity, specificity, AUC, and RR did not differ significantly between ELISA and non-ELISA methods. CONCLUSIONS: Generally, the non-ELISAs showed better performance for selected analytes, particularly aCL IgG, while ELISAs demonstrated greater consistency. Although agreement between methods varied for specific antiphospholipid antibodies, overall diagnostic accuracy was comparable when the complete antiphospholipid antibody profile was assessed. Clinicians should consider these factors to ensure accurate and reliable diagnostics.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers