B-164 Multicenter evaluation of the new Access HBsAg assay with highly sensitive HBsAg detection on the DxI 9000 Access Immunoassay Analyzer in the United States

Jennifer Kalamatas, Robert Christenson, Fred Apple, Saeed Jortani, Florian Esnault, Marc Turini, Danielle Maher

Key points

From the paper's conclusion
  • The Access assay also offered an analytical sensitivity (limit of detection) for the HBsAg WHO I.S. of 0.024 IU/mL.

Abstract

Background Beckman Coulter developed fully automated Access HBsAg assay for the qualitative detection of hepatitis B virus surface Antigen (HBsAg) in human serum or plasma for the new DxI 9000 Immunoassay Analyzer. Evaluation of the Access HBsAg assay assessed imprecision, analytical sensitivity for seroconversion panels and for the Third WHO I.S. (NIBSC code 12/226), and detection of HBsAg from different viral genotypes and mutants. Positive and negative percent agreements of these assays were also assessed compared to an FDA approved reference assay. Methods Imprecision was assessed during a 20-day study performed on the DxI 9000 Immunoassay Analyzer. A ten-member panel of patient samples, including serum and plasma samples were assayed using three lots of Access HBsAg reagent kits, on one instrument. Analytical sensitivity for seroconversion panels was conducted using thirty commercially available seroconversion panels assessed on Access HBsAg (3 lots) and FDA approved comparator assay. Analytical sensitivity for WHO was performed on dilutions of HBsAg 3rd WHO I.S. (NIBSC code 12/226) prepared in HBsAg negative human serum and plasma, ranging from 0.02 to 0.024 IU/mL. Genotype detection was evaluated using a total of 24 panel members (A (n=4), B (n=3), C (n=4), D (n=4), E (n=2), F (n=3), G (n=1) and H (n=3)). A total of 30 (10 native and 20 recombinant) HBsAg mutant samples, containing defined mutations between amino acid 100 and 170 and representing the most common HBsAg mutants reported in literature, were also tested. Negative and positive percent agreements were determined for the Access assay using the FDA approved reference assay as comparator by testing samples from patients classified as increased risk for hepatitis due to lifestyle, behavior, occupation, or known exposure events, or individuals with signs and symptoms of hepatitis. Results Imprecision results showed a maximum repeatability (within-run) of 1.7% CV for positive patient samples and 0.00 S/CO SD for negative patient samples. The maximum patient sample within-laboratory imprecision (total) obtained was 6.9% CV on positive samples and 0.01 S/CO SD on negative samples. The Access assay was later for three panels but equivalent on 27 panels in terms of detection day of first positive bleed compared to FDA approved reference assay. The difference in the mean first day of detection across the 30 seroconversion panels was less than one day (0.7 days) between the two assays. The Access assay also offered an analytical sensitivity (limit of detection) for the HBsAg WHO I.S. of 0.024 IU/mL. All 24 genotype samples and all 30 mutants tested were reactive by the Access HBsAg assay. Clinical validation study conducted at three U.S. clinical sites located in Minneapolis, Louisville or Baltimore demonstrated 99.5% PPA between Access and reference assay (n= 417; 95% CI: 98.3-99.9%) and 99.7% NPA (n=3,197; 95% CI: 99.5-99.9%) between Access HBsAg and reference assay. Conclusion The Access HBsAg assay demonstrated analytical and clinical performance equivalent to US marketed HBsAg assays. It also provides features regarding its low samples volume requirements, robustness, and processes that do not require systematic retesting nor confirmation assay for high positive samples.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers