B-092 Evaluation of potential analytical interferents in samples using the Abbott Alinity i CA19-9XR immunoassay

Roma Kulkarni, Megan Lau, Ana Shulla-Mesi, John Jaraczewski, Lori Sokoll, Gerard Davis, Gavin Cloherty

Key points

From the paper's conclusion
  • This study was undertaken to identify and characterize analytical interferents potentially present in samples with larger CA19 9 concentration differences measured by assays from two major IVD manufacturers (Abbott and Roche).

Abstract

Background Abbott’s CA19-9XR assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma on the Alinity i analyzer. The Alinity i CA19-9XR assay is used as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods. Despite a lack of supporting data, CA 19-9 values are used for other purposes (e.g., screening) in some geographic areas. CA19-9 results across different assay platforms are known to show considerable variability due to differences in assay methods, reagent specificity, lack of standardization and, potentially, sensitivity to various analytical interferences. This study was undertaken to identify and characterize analytical interferents potentially present in samples with larger CA19 9 concentration differences measured by assays from two major IVD manufacturers (Abbott and Roche). Methods A total of 1,882 consecutive remnant serum specimens submitted for CA19 9 testing at Johns Hopkins Medical Institutions (JHMI) were evaluated. Samples were categorized into four groups by comparing results from the Roche Elecsys CA19-9 assay with those from the Abbott Alinity i CA19-9XR assay, using the respective manufacturer-defined reference ranges (Table 1). Sixty of the specimens demonstrating the largest percentage differences between the two assays were selected for further characterization. The 60 specimens were equally divided between each manufacturer having the higher value. A testing algorithm was completed by initially confirming the presence of 1116-NS-19-9 reactive determinants by treating the samples with neuraminidase, an enzyme that removes sialic acid residues from the CA19 9 antigen. A significant reduction in the CA19-9 signal following neuraminidase treatment confirms the presence of the CA19-9 analyte in the sample. In parallel, samples were also treated with polyethylene glycol (PEG) to precipitate high molecular weight proteins and/or antibodies that could contribute to assay interference. Results from these treatments were used to determine if additional characterization (e.g., Heterophilic Blocking Tubes (HBT)) was warranted. Results The overall concordance between the Abbott and Roche assays using the respective manufacturer-defined reference ranges was 90.3%, which supports previously published findings. Neuraminidase and PEG treatments confirmed the integrity (or lack of analytical interference) of all Alinity i CA19-9 values from the 60 discrepant samples selected in this study, indicating other reasons (differences in assay methods, reagent specificity and/or lack of standardization) for differences between the Abbott and Roche assays. Conclusion Abbott and Roche results in this study compared well despite differences in assay methods, reagent specificity and/or lack of standardization. Across 1,882 samples, no specimens were found to contain an interferent that resulted in clinical discordance (with respect to reference ranges) between the Abbott and Roche platforms. Until tumor markers, such as CA19-9 are standardized, results between manufacturers’ assays should be interpreted with discretion and in conjunction with other clinical variables.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers