Improving tumor marker comparability: A case series on harmonizing carbohydrate antigen 19‐9 and carbohydrate antigen 15‐3 measurements across diverse measurement systems
Jia Li, Xiaoying Lou, Zongdan Wang, Mengdi Wu, Jing Zhang, Huishuang Gao, Meng Li, Bing Wu,
Key points
From the paper's conclusion- We evaluated the comparability of carbohydrate antigen 19‐9 (CA19‐9) and carbohydrate antigen 15‐3 (CA15‐3) across seven contemporary measurement systems—Abbott Alinity I, Autobio A2000PLUS, Beckman Access Dxl 800, Mindray CL6000i, Roche Cobas E801, Snibe Maglumi X8, and Sysmex HISCL‐5000 using 464 native serum samples for CA19‐9 and 499 for CA15‐3.
Abstract
Serum tumor markers play a critical role in the early screening, treatment monitoring, and prognostic assessment of malignant tumors, yet poor comparability across different measurement systems frequently undermines their clinical value. We evaluated the comparability of carbohydrate antigen 19‐9 (CA19‐9) and carbohydrate antigen 15‐3 (CA15‐3) across seven contemporary measurement systems—Abbott Alinity I, Autobio A2000PLUS, Beckman Access Dxl 800, Mindray CL6000i, Roche Cobas E801, Snibe Maglumi X8, and Sysmex HISCL‐5000 using 464 native serum samples for CA19‐9 and 499 for CA15‐3. Spearman correlations were higher for CA15‑3 ( r : 0.927‒0.997) than for CA19‑9 ( r : 0.777‒0.985). With the all‐method median as a pragmatic comparator, Passing‒Bablok regression and difference plots were generated for each system. Against analytical performance specifications derived from biological variation, all systems except Abbott Alinity (22.5%) met the minimum bias criterion of ±21.6% for CA19‐9, whereas for CA15‐3 only Abbott Alinity (‒10.159%), Roche Cobas (13.01%), and Snibe Maglumi (−3.923%) fell within the minimum bias criterion of ±13.9%. CA19‐9 and CA15‐3 concentrations were recalibrated using three mathematical approaches: principal component analysis and two nonlinear regression models. After recalibration, the variability of CA19‐9 and CA15‐3 values decreased significantly. Our findings indicate that CA19‐9 and CA15‐3 results are not directly interchangeable across different systems, which poses a real‐world challenge for longitudinal patient follow‐up. However, mathematical models derived from individual patient serum samples can effectively reduce system‐related discrepancies, thereby enabling more consistent serial interpretations and supporting more informed clinical decisions in routine practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-079 Best Practices for Tumor Marker Re-baselining: Practical Insights from CA 19-9 Assay ConversionClinical Chemistry · 2026
- Comparative Analytical Performance and Clinical Agreement of Chemiluminescence Microparticle Immunoassay and Electrochemiluminescence Immunoassay for CA19-9: A Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026
- Clinical significance of inter-assay discrepancies in serum CA19-9 values: A comparison between CLIA and ECLIA.Annals of clinical biochemistry · 2026
- Harmonization and standardization of tumor markers and liquid biopsies in China: A short perspective.Tumour biology : the journal of the International Society for Oncodevelopmental Biology and Medicine · 2026
- Verification of a previously obtained EQA based MUC1/CA15-3 harmonisation procedure.Tumour biology : the journal of the International Society for Oncodevelopmental Biology and Medicine · 2026
- Beyond proficiency testing: Leveraging commutable External Quality Assessment data to harmonize tumor markers and refine medical decision points.Tumour biology : the journal of the International Society for Oncodevelopmental Biology and Medicine · 2026