Clinical Chemistry

B-079 Best Practices for Tumor Marker Re-baselining: Practical Insights from CA 19-9 Assay Conversion

Anil Kiran Chokkalla, Christopher Koch, Arun Rathinam, Mark Gitau, Jennifer Rider, Kim Bennett

Key points

From the paper's conclusion
  • These findings are consistent with prior external quality assessment data, which also reported a substantial positive bias (∼77%) for the Abbott method.

Abstract

Background Assay conversion for tumor markers requires re-baselining of patient results, particularly when transitioning to a new instrument platform, addressing reagent lot-to-lot bias, or bringing previously outsourced testing in-house. Because most tumor markers lack harmonization across analytical methods, implementing a structured re-baselining protocol is critical to ensure safe and reliable monitoring of treatment response and cancer recurrence. There is limited evidence-based guidance on re-baselining protocols, including recommended time intervals and the analytical and clinical criteria required for interpretation. The goal of this study is to develop a best-practice framework for tumor marker re-baselining, using CA 19-9 assay conversion within our health system as an example. Methods To internalize the CA 19-9 testing at our health system on the Abbott Alinity platform from reference laboratory, which utilizes Siemens ADVIA Centaur platform, we devised a strategy for re-baselining patient results. Re-baselining was conducted for a period of 3 months , where each specimen received by our laboratory is tested in parallel using both Abbott and Siemens methods. To support consistent specimen handling, we created an orderable panel that enabled simultaneous accessioning to in-house and send-out test. An automated comment was included on each result to inform providers about the ongoing re-baselining process. At the end of the re-baselining period, Abbott and Siemens results were analyzed using Deming regression, allowing estimation of both constant and proportional bias. Results A total of 158 specimens were included in the re-baselining study. Method comparison demonstrated significantly higher CA 19-9 values on Abbott method compared to Siemens method. Deming regression analysis yielded a slope of 2.46 (95% CI: 2.33 to 2.59), intercept of -49.9 U/mL (95% CI: -186.6 to 86.8) and a correlation coefficient of 0.94 (Figure 1). Mean bias between methods was ∼134%, with increasing variation at higher concentrations. These findings are consistent with prior external quality assessment data, which also reported a substantial positive bias (∼77%) for the Abbott method. Conclusion Our re-baselining strategy incorporates parallel testing, optimal design of test build, targeted provider communication, and clinical correlation to ensure safe monitoring of tumor markers during assay conversion (Figure 1). The substantial variability in the intercept across the analytical measurement range demonstrates that constant bias is not stable, underscoring the need to establish new patient-specific baselines rather than relying on mathematical back-calculation of historical CA 19-9 results.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers