B-084 Comprehensive Individualized Quality Control Plan (IQCP) Implementation and Analytical Verification of BCR-ABL1 p210 Transcript Quantification on the GeneXpert Platform
Shivaprasad H Sathyanarayana, Alejandro Garcia Ruiz, Patrick M Shannon, Micheal A Johnston, Jeremiah X Karrs, Liam L Donnelly, Laura J Tafe, Joel A Lefferts
Key points
From the paper's conclusion- Overall, four QC failures were observed; however, all results remained within ± 2 SD of the established means.
Abstract
Background The BCR::ABL1 fusion is a critical diagnostic marker in Chronic Myeloid Leukemia (CML), and quantitative measurement of BCR::ABL1 transcripts by RT-PCR is essential for therapeutic monitoring. The Xpert® BCR-ABL Ultra (Cepheid) is an automated, cartridge-based system that integrates sample preparation and RT-PCR with fluorescent detection, targeting the predominant p210 transcript. In CLIA-certified laboratories, an Individualized Quality Control Plan (IQCP) is required when quality control (QC) materials are not tested daily with patient specimens. We developed and implemented an IQCP for the Xpert assay in 2022 and repeated the process following installation of a new instrument in 2026. This study presents a comparative analysis of IQCP performance between the two instruments, along with over 36 months of weekly QC monitoring data using commercially available external controls. Methods The Xpert BCR-ABL1 IS panel (MMQCI, C130) that includes reference materials with BCR::ABL1 concentrations of 0%, 0.0032%, 0.01%, 0.1%, 1%, and 10% on the International Scale (IS) was used in this study. For IQCP development and analytical verification, the 0.01% (low positive control) and 1% (high positive control) panel members were tested on eight consecutive daily runs in our new instrument. Expected MR ranges were established using the mean molecular response (MR) value ±2 standard deviations (SD). For ongoing quality control, the 0% (negative), 0.01% (low positive), and 1% (high positive) controls were tested weekly with clinical testing for approximately thirty-six weeks. QC monitoring also included evaluation of significant shifts in control values associated with reagent cartridge lot changes or control material lot variations. Results Eight consecutive daily runs of the positive controls yielded mean minimal residual (MR) values of 4.12 ± 0.16 for the low positive control and 1.83 ± 0.12 for the high positive control. In comparison, the corresponding 2022 values were 4.09 ± 0.29 (low positive) and 1.87 ± 0.20 (high positive), demonstrating minimal variation between the two evaluations. Based on these data, the established mean ± 2 standard deviation (SD) acceptance ranges were 3.80-4.44 (2026) versus 3.51-4.68 (2022) for the low positive control, and 1.59-2.07 (2026) versus 1.48-2.26 (2022) for the high positive control. Ongoing weekly QC monitoring over 36 months of patient testing, encompassing 10 lots of the Xpert assay kit and 5 lots of the Xpert BCR-ABL1 IS panel, demonstrated minimal variability (low positive mean 4.10 ± 0.24; high positive mean 1.86 ± 0.13). Overall, four QC failures were observed; however, all results remained within ± 2 SD of the established means. Minor shifts in positive control values were noted following assay lot changes but did not exceed the predefined mean MR ± 2 SD acceptance criteria Conclusion Our laboratory implemented an IQCP for the Xpert BCR-ABL assay with weekly QC monitoring. Longitudinal QC analysis demonstrates sustained assay performance, minimal variability, and no significant deviations over 36 months of patient testing. Analytical verification of the new GeneXpert instrument using the Xpert® BCR-ABL Ultra demonstrated performance comparable to the previous instrument, with faster turnaround time and reduced hands-on workflow.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a Multiplex PCR to Detect the BCR-ABL1 Transcripts in Chronic Myeloid Leukaemia Patients in Indonesia.Cancer management and research · 2026
- Next-generation digital polymerase chain reaction assay for BCR::ABL1 in chronic myeloid leukemia: comprehensive validation for standardized measurable residual disease detection.Blood research · 2026
- B-405 Estimation of reference change value for the cyclosporine assay and implementation of delta check auto verificationClinical Chemistry · 2026
- B-252 Prospective real-time implementation and clinical validation of patient-based real-time quality control using average of normals in a tertiary hospital laboratorClinical Chemistry · 2026
- Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection SystemsMicroorganisms · 2026
- A-307 Determining risk-based internal quality control frequency for biochemical testing using Sigma metrics and the QC Constellation toolClinical Chemistry · 2026