B-405 Estimation of reference change value for the cyclosporine assay and implementation of delta check auto verification
Hannah Mbatia, Tabitha Wilson, Elisha Roberts, Gloria Vashon, Harris Neil, Bremansu Osa-Andrews, Maximo Marin
Key points
From the paper's conclusion- This estimated RCV was subsequently adopted as the delta check reference change value for the cyclosporine assay.
Abstract
Background Cyclosporine, a transplant immunosuppressant medication, shows substantial intra and inter-individual biological variability due to differences in absorption, bioavailability, and metabolism. Therapeutic drug monitoring is routinely employed to guide dosing, maintain appropriate drug levels, and prevent organ rejection and nephrotoxicity. Labor-intensive pretreatment, including multiple pipetting steps, intermittently introduces random errors into many whole blood-based cyclosporin assays. These analytical challenges, combined with the drug’s inherent biological variability and relatively wide therapeutic window, make detection of errors difficult. The Reference Change Value (RCV) provides a threshold beyond which the difference between two consecutive results of an individual patient is either clinically critical or attributed to error. Derived from analytical variation (CVA) and within-individual biological variation (CVI), In the present study, RCV was optimized as an auto verification delta check for cyclosporin, to enhance the detection and reduction of testing errors. Methods The study utilized serial cyclosporine results of 12 patients extracted from the laboratory information system (LIS) over a 60-day period. Cyclosporine levels were analyzed using the Architect Cyclosporine Whole Blood kit designed for use with the Architect 2000i analyzer. Only results within the acceptable therapeutic window of 0–700 ng/mL were included in the data analysis. Excel spreadsheets were used to analyze data. The estimated RCV for cyclosporin was calculated using the formula: RCV = Z x 21/2 (CVA2 + CVI2) ½, Z represents the number of SD from the mean in the standard normal distribution (2.58 % probability for 99%). CVA - analytical coefficient of variation and CVI - biological coefficient of variation. CVA was obtained from the monthly QC data. The CVI was estimated using the mean and standard deviations of a set of serial patient measurements. The product of the %RCV and the average cyclosporin result of the patients computed the estimated RCV. Results The mean cyclosporin results of individual patients ranged from 81.53 ng/mL to 239.44 ng/mL with an overall concentration average of 161.86 ng/mL. The corresponding standard deviations ranged from 45.34 ng/mL to 139.40 ng/mL. The mean CVI was calculated to be 51.98% and the CVA was 28.8%. Combining the mean CVI, the CVA and the z-score in the RCV formula generated an estimated RCV of 216 %. Using the average concentration of 161.86 ng/mL, this corresponded to an absolute RCV of 349.61 ng/mL. This estimated RCV was subsequently adopted as the delta check reference change value for the cyclosporine assay. Delta check auto-verification was successfully validated in the LIS through repeated dry testing and thereafter implemented into the routine testing workflow. Conclusion Here, we report the application of estimated RCV as a delta-check auto verification value at our core laboratory to assess the total allowable analytical and biological variation for the cyclosporine assay. Patient results that exceed the established RCV delta threshold are withheld for further investigation, allowing the laboratory to identify erroneous results. Taken together, this study demonstrates an effective quality improvement that enhances patient safety for an error-prone but critical test, cyclosporin.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Long-Term Monitoring of Everolimus Concentrations in Routine Clinical Practice: Evaluation of Patient Percentiles as Quality Indicator and Impact of Analytical Method Change.Pharmaceuticals (Basel, Switzerland) · 2026
- B-414 Analytical Performance Evaluation of a Cyclosporine Assay Under Development for the Alinity i PlatformClinical Chemistry · 2026
- Investigating the Variable Component of the Systematic Error, a Neglected Error Parameter: Theoretical Reevaluation StudyJMIRx Med · 2026
- Long-term analytical drift of serum creatinine and its association with CKD staging consistency based on eGFR: a retrospective study.Scandinavian journal of clinical and laboratory investigation · 2026
- Circular validation in diagnostic research: re-examination of a widely cited study on primary hemostasis screening.Frontiers in medicine · 2026
- Monitoring Long-term Tacrolimus Exposure: Metrics, Trajectories, and Associations with Outcomes after Kidney Transplantation in the Wisconsin Allograft Recipient Database (WisARD).American journal of nephrology · 2026