B-050 Implementation and Performance Evaluation of Automated Autoverification for Tumor Marker Results in a High-Complexity Oncology Laboratory

Shirley Rojas Sanchez, Rommy Pizarro

Key points

From the paper's conclusion
  • Conclusion Automatic autoverification was applicable to a relevant proportion of tumor marker results, particularly CA15-3, PSA, and CA125, demonstrating that selected analytes with stable biological behavior can be safely automated without compromising analytical quality.

Abstract

Background Automatic result autoverification is a key strategy for optimizing clinical laboratory workflow by improving efficiency, standardization, and turnaround time. In high-complexity oncologic institutions, where tumor markers guide therapeutic decision-making, treatment response monitoring, prognostic assessment, and changes in clinical management, result validity and patient safety are critical. Oncologic populations present particular challenges, including high biological variability, close longitudinal follow-up, and results frequently near clinical decision thresholds, which complicate the implementation of autoverification systems. In this context, automated validation rules must be aligned with international standards and tailored to each analyte. The aim of this study was to evaluate the implementation of automatic autoverification of tumor markers in the immunology laboratory of a tertiary oncologic institute in Lima, Peru. Methods A two-year retrospective observational descriptive study was conducted. A total of 146,854 tumor marker results processed in the immunology laboratory were analyzed. Analytical testing was performed using the Atellica® IM 1600 platform (Siemens Healthineers) integrated with Atellica Data Manager (ADM) middleware. Autoverification was configured in ADM using predefined rules that included analyte-specific reference intervals, delta checks based on the relative change value (RCV) calculated according to CLSI EP33 guidelines, evaluation of internal quality control performance, and verification of instrument alarms and flags. Results were classified as automatically autoverified or manually validated by a laboratory professional. Results Of the 146,854 results, 42,914 (29%) were released through automatic autoverification, while 103,940 (71%) required manual validation. The analytes with the highest autoverification rates were CA15-3 (44%), PSA (36%), and CA125 (36%), indicating that tumor markers with greater biological stability and well-defined clinical ranges allow higher levels of automation. In contrast, CA19-9 (25%), CEA (26%), and free PSA (10%) showed lower autoverification rates, with a predominance of manual validation, likely due to greater biological variability and results close to clinical decision thresholds requiring correlation with patient status. No autoverified results were observed for thyroglobulin or anti-thyroglobulin antibodies (aTgII), used in thyroid cancer follow-up, owing to their high biological variability and need for clinical interpretation. Similarly, human chorionic gonadotropin (hCG) and alpha-fetoprotein (AFP) were not autoverified (100% manual validation) because of their high clinical impact and the risk of analytical interferences such as the hook effect, necessitating review of clinical history. Conclusion Automatic autoverification was applicable to a relevant proportion of tumor marker results, particularly CA15-3, PSA, and CA125, demonstrating that selected analytes with stable biological behavior can be safely automated without compromising analytical quality. Lower autoverification rates for CEA, CA19-9, and free PSA highlight the need for clinical correlation and careful monitoring of biological variability in oncologic populations. The absence of autoverification for thyroglobulin, hCG, and AFP underscores the importance of manual validation for analytes with high variability or interference risk that may impact critical therapeutic decisions. Optimization of autoverification requires analyte-specific rules and periodic review of reference intervals and delta checks, with criteria adapted to the characteristics of oncologic patients.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-050 Implementation and Performance Evaluation of Automated Autoverification for Tumor Marker Results in a High-Complexity Oncology Laboratory — Science Skim