A-255 Presumptive identification of rare hemoglobin variants by CAPILLARYS 3 Hemoglobin(e) assay.
Matthew Wagner, Wyatt Cramblet, Aigars Brants
Key points
From the paper's conclusion- Conclusion The CAPILLARYS 3 Hemoglobin(e) assay, when conducted with confirmatory testing provided sufficient information to correctly narrow down the list of potential variants to a handful of possibilities containing the true result.
Abstract
Background the CAPILLARYS 3 Hemoglobin(e) assay separates abnormal hemoglobins from the A, A2, and F (if present) fractions by their differing electrophoretic mobility. This separation places the hemoglobin variants into distinct regions (zones) on the electropherogram for distinguishing common variants from one another, but the large number of variants that have been identified in the literature means that each region hosts numerous rare variants as well. Distinguishing these from one another requires careful analysis by profile type (number/relative position of variant peaks) position (zone and x-position of the individual peak), and percentage (relative percent expression). Even then, confirmation by an alternate assay type is required. Common confirmatory techniques include HPLC (High performance Liquid Chromatography, separating hemoglobin species by differing affinity to a cation-exchange column), sickling tests (for confirmation of Hb S presence), and acid gel electrophoresis (separation by electrophoretic mobility on a low pH gel media). Consideration of results from both capillary and the confirmatory assay provides presumptive identification of rare variants, but only genetic analysis can provide a conclusive result. Methods The Result Interpretation Escalation (RIE) group is a service provided for users of the Sebia Hemoglobin(e) assay, providing additional interpretation assistance on distinguishing rare hemoglobin variants from one another based on an internal library of published sources and internal data (both on Capillary and other confirmatory testing). Variant profiles, positions, and percent representation are considered in sequence, and the potential variants indicated are then considered in relation to the confirmatory results. The examples presented here are cases submitted to RIE, demonstrating the algorithm which is applied to these results, the presumptive identification reached, and the results as eventually returned from genetic testing for those specific cases. Individual cases were chosen to demonstrate the degree of certainty capable through comparison of the CAPILLARYS 3 Hemoglobin(e) result to confirmatory methods. Results The CAPILLARYS Hemoglobin(e) assay and confirmatory assay results for selected cases were employed to present a list of possible variants. Subsequent genetic testing confirmed the identity of the variant present, and in each case the final confirmation was found to be one of the listed potential variants. Variants included in these cases were Hb Canuts, Hb Chesapeake, Hb Russ, Hb Wenchang-Wuming, Hb Nigeria, Hb Old Dominion, Hb Presbyterian, Hb Hope, and Hb Q-Thailand. Conclusion The CAPILLARYS 3 Hemoglobin(e) assay, when conducted with confirmatory testing provided sufficient information to correctly narrow down the list of potential variants to a handful of possibilities containing the true result. These cases demonstrate the efficacy of both the technique and the analysis algorithm when distinguishing between rare and common hemoglobin variants.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detection of a Silent Hemoglobin Variant During HbA1c Monitoring: the Crucial Role of Capillary Electrophoresis.Clinical laboratory · 2026
- The Efficiency of Capillarys 2 Flex Piercing, Premier Resolution-HPLC, and VARIANT II-HPLC for Hemoglobin Constant Spring Detection.Hemoglobin · 2026
- Identification of Clinically Relevant Hb Variants During HbA1c Testing Using the ARKRAY HA-8180V in Fast Mode: Reliable HbE Detection Based on Retention Time.Annals of laboratory medicine · 2026
- B-299 Analytical performance, variant detection and result reportability of the CAPILLARYS 3 TERA HbA1c assay relative to the Siemens Atellica CH930Clinical Chemistry · 2026
- 70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting IssuesAmerican Journal of Clinical Pathology · 2026
- Evaluation of Analytical Agreement Between High-Performance Liquid Chromatography and Capillary Electrophoresis for Haemoglobin A2 Quantification: A Meta-Analysis.The Malaysian journal of medical sciences : MJMS · 2026