Evaluation of Analytical Agreement Between High-Performance Liquid Chromatography and Capillary Electrophoresis for Haemoglobin A2 Quantification: A Meta-Analysis.
Yongzhong Wei, Mingze Gu, Yuxin Su, Jiayun Xi, Yuqing Ren, Zhongshuang Hu
Institute of Biomaterials and Biomedicine, School of Food and Pharmacy, Shanghai Zhongqiao Vocational and Technical University, Shanghai, People's Republic of China.
Key points
From the paper's conclusion- HPLC and CE demonstrate systematic but clinically relevant differences in HbA2 measurement, particularly near diagnostic thresholds. HbA2 measurement results obtained from HPLC and CE should not be used interchangeably, and method-specific interpretation and harmonisation efforts are warranted.
Abstract
BACKGROUND: Haemoglobin A2 (HbA2) measurement is essential for diagnosing β-thalassemia carriers, yet even small analytical biases may affect clinical interpretation near the diagnostic threshold of 3.5%. High-performance liquid chromatography (HPLC) and capillary electrophoresis (CE) are widely used, but their agreement remains uncertain. The study aimed to systematically evaluate the measurement bias between HPLC and CE in HbA2 quantification. METHODS: A systematic review and meta-analysis were conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines. Eligible studies directly compared HbA2 values measured by HPLC and CE in the same samples. Data on mean bias, standard deviation (SD), and limits of agreement (LoA) were extracted. A random-effects model was used to pool mean bias, and subgroup analyses were performed for normal individuals, β-thalassemia carriers, and haemoglobin variant carriers. Sensitivity analysis was conducted to assess robustness. RESULTS: Ten studies with 1,881 paired measurements were included. After excluding datasets with clear co-elution interference, mainly haemoglobin E (HbE) and haemoglobin Hope (Hb Hope), the pooled mean bias was 0.16% (95% confidence interval [CI]: 0.03, 0.29), with high heterogeneity (I 2 = 99.5). Subgroup analysis showed a mean bias of 0.37% (95% CI: 0.24, 0.50) in healthy individuals and 0.39% (95% CI: 0.20, 0.59) in β-thalassemia carriers. Sensitivity analysis confirmed the stability of the results. CONCLUSION: HPLC and CE demonstrate systematic but clinically relevant differences in HbA2 measurement, particularly near diagnostic thresholds. HbA2 measurement results obtained from HPLC and CE should not be used interchangeably, and method-specific interpretation and harmonisation efforts are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating HbA1c measurement in haemoglobinopathy patients: Is transition beyond HPLC warranted for improved clinical accuracy?Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine MethodsClinical Chemistry · 2026
- 70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting IssuesAmerican Journal of Clinical Pathology · 2026
- B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic VariantsClinical Chemistry · 2026
- A-131 HbA1c Result Patterns by Capillary Electrophoresis in Primary-Use and Reference-Use Laboratory PopulationsClinical Chemistry · 2026
- Impact of ISO/TS 20914-based measurement uncertainty on HbA1c diagnostic classification: A real-world comparison of two analytical systems.Clinica chimica acta; international journal of clinical chemistry · 2026