Identification of Clinically Relevant Hb Variants During HbA1c Testing Using the ARKRAY HA-8180V in Fast Mode: Reliable HbE Detection Based on Retention Time.

Monthon Mayoon, Sitthichai Panyasai

一句话结论 · In one sentence

HbE and thalassemia variants can interfere with HbA1c quantification. ARKRAY HA-8180V FM chromatogram review allows reliable detection of HbE and prevents misinterpretation of HbA1c results.

原始摘要(英文原文)· Original abstract
BACKGROUND: Hb variants are often incidentally detected during HbA1c testing and can cause interference. We evaluated the ability of the ARKRAY HA-8180V HPLC system operating in fast mode (FM) to identify Hb variants, particularly HbE, in samples submitted for HbA1c testing based on characteristic retention times (RTs). METHODS: Chromatograms of samples tested for HbA1c were analyzed; those showing predefined abnormal chromatographic patterns suggestive of Hb variants were further evaluated using capillary electrophoresis (CE) and β- and α-globin gene sequencing. RTs were recorded. RESULTS: Of 4,888 chromatograms analyzed, 588 (12.0%) were abnormal. Among these, 564 (95.9%) showed an abnormal peak at an RT of 23-25 s and were confirmed to contain HbE via CE. Identified Hb phenotypes included EA (N=526), compound heterozygotes [Constant Spring (CS) EA, N=2; CSEE, N=1], EE (N=27), EFA (N=3), and EF (N=5). Other variants included Hb CSA2ABart'sH, A2FA, A2FABart's, and A2F. Rare variants, Hb Hope (N=12) and Hb Korle-Bu (N=1), and two uncharacterized variants were also detected. Hb Hope interfered with HbA1c measurements, and all CSA2ABart'sH samples produced falsely low HbA1c values. Artificially decreased HbA1c was detected in 21/27 (77.8%) HbEE samples. ARKRAY HA-8180V FM analysis yielded reproducible RTs for HbE, achieving 100% sensitivity, specificity, and accuracy in the validation cohort. CONCLUSIONS: HbE and thalassemia variants can interfere with HbA1c quantification. ARKRAY HA-8180V FM chromatogram review allows reliable detection of HbE and prevents misinterpretation of HbA1c results.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Identification of Clinically Relevant Hb Variants During HbA1c Testing Using the ARKRAY HA-8180V in Fast Mode: Reliable HbE Detection Based on Retention Time. — 科研速览 Science Skim