Methemoglobin Causes Collapsed WDF Scattergrams and a Low-SFL Mature RBC Population in RET Scattergrams on the XR-Series Automated Hematology Analyzer.
Takahiro Kaido, Yuka Takezawa, Akari Miyazaki, Yuichiro Ide, Nau Ishimine, Yoko Usami
Department of Laboratory Medicine, Shinshu University Hospital, Matsumoto, Nagano, Japan.
Key points
From the paper's conclusion- MetHb produces characteristic abnormalities in WDF and RET scattergrams on the XR-Series analyzer. These findings may provide useful clues for the incidental detection of unsuspected methemoglobinemia during routine CBC testing.
Abstract
INTRODUCTION: Methemoglobinemia is characterized by increased methemoglobin (MetHb), which impairs oxygen delivery and may cause clinically significant hypoxia. Because CBC testing is more routinely performed than blood gas analysis, abnormalities detected during CBC testing may provide useful clues to unsuspected methemoglobinemia. We investigated the effects of MetHb on scattergrams and optical indices generated by the XR-Series automated hematology analyzer. METHODS: MetHb-containing samples were prepared from five healthy donors using sodium nitrite-treated red blood cells and analyzed using the XR-9000. WDF, RET, and PLT-F scattergrams and WDF sensitivity indices were evaluated. Reference intervals were established using data from 812 individuals in the reference population. Potassium cyanide (KCN) treatment was performed on MetHb-containing samples from all five donors. Clinical cases were examined. RESULTS: Increasing MetHb concentrations reduced side fluorescence (SFL) and produced a collapsed WDF scattergram pattern. SFL reductions were more pronounced in lymphocytes and monocytes than in neutrophils, and LY-Y and MO-Y fell below their lower reference limits even at low MetHb concentrations. MetHb induced a distinct low-SFL population of mature red blood cells in RET and PLT-F scattergrams, distinguishable even at low MetHb concentrations in the RET channel. KCN treatment restored WDF and RET scattergrams to patterns similar to those of the normal control in all five donors. Similar abnormalities were observed in clinical methemoglobinemia. CONCLUSION: MetHb produces characteristic abnormalities in WDF and RET scattergrams on the XR-Series analyzer. These findings may provide useful clues for the incidental detection of unsuspected methemoglobinemia during routine CBC testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-204 Chemical modification of the hemolysis index with methylene blue on an automated chemistry analyzer: The holy grail?Clinical Chemistry · 2026
- Evaluation of Reticulocyte Channel on Sysmex XN ‐10 for Resolving Cold Agglutination in Samples With High Prevalence of Thalassemic HemoglobinopathiesInternational Journal of Laboratory Hematology · 2026
- Hematologic follow-up of metamizole-associated Heinz body anemia in dogs.Frontiers in veterinary science · 2026
- Routine CBC-based machine learning for individualized correction of lipemia-induced hemoglobin bias.Frontiers in medicine · 2026
- Analytical discrepancy in platelet counting in MYH9-related disease: a case report.Laboratory medicine · 2026
- CD71-based immature RBC (iRBC) analysis in peripheral blood has a better correlation with leukocyte PNH clone size and improves the delineation of different PNH clone types.Cytometry. Part B, Clinical cytometry · 2026