A-192 Evaluating the potential impact of leached preservative on chemical urinalysis in residual midstream urine specimens

Leah Brown, Michelle Seaton, Gurjot Dhaliwal, Nicole White-Al Habeeb, Lee Goneau

Key points

From the paper's conclusion
  • This study evaluated whether manufacturer-recommended (5-second) or prolonged (60-second, worst-case) preservative exposure affects chemical urinalysis results on the Clinitek Novus (Siemens Healthineers).

Abstract

Background UriSponge® (Copan Diagnostics) is a urine collection and preservation system containing sodium propionate and potassium sorbate–impregnated sponges designed to stabilize specimens for culture. Manufacturer instructions specify a 5-second dip of the sponge applicator into the urine sample. This sampling process introduces preservatives into residual urine, raising concern for potential impact on downstream chemical urinalysis. This study evaluated whether manufacturer-recommended (5-second) or prolonged (60-second, worst-case) preservative exposure affects chemical urinalysis results on the Clinitek Novus (Siemens Healthineers). Methods In the primary evaluation, 426 urine specimens were analyzed before and after manufacturer-recommended 5-second UriSponge dipping. Specific gravity (SG) was assessed by two-instrument comparison using a total allowable error (TEa) of 0.003. Urine pH agreement was evaluated categorically using predefined criteria of at least 95% within ±1 pad, no more than 5% with 2-pad or greater differences, and absence of systematic directional bias. Ten semiquantitative parameters (bilirubin, blood, glucose, ketones, leukocyte esterase, nitrite, protein, urobilinogen, colour, clarity) were also compared. For extended exposure assessment, 65 specimens were aliquoted into control, 5-second dip, and 60-second (worst-case) dip groups and analyzed after 24-hour storage under refrigerated (n=34) or room temperature (n=31) conditions. Results In the primary study, SG at baseline (0 minutes) and 30 minutes post–5-second dip were equivalent within a TEa of 0.003 across 426 specimens (range 1.002–1.046), with 99.8% within allowable error and excellent agreement by Deming regression (R=0.9977; slope 0.998; intercept 0.0024; SEE 0.0005; mean Error Index 0.00). For pH, 85.0% showed exact agreement, 14.1% differed by ±1 pad, and 0.9% by 2 pads or more (99.1% within ±1 pad). Semiquantitative analyte agreement ranged from 90.6% to 99.1%. After 24-hour storage (n=65), SG remained equivalent to 24-hour controls for both 5- and 60-second dips (98.5% within TEa), with strong agreement (5s: R=0.9935, slope 0.996; 60s: R=0.9943, slope 0.987) and minimal mean bias. For pH, no systematic bias was observed with 5-second exposure; however, 60-second exposure demonstrated increased variability at room temperature. Qualitative agreement after 24 hours ranged from 89.2% to 100% across analytes, with leukocyte esterase showing the lowest agreement (89.2%). Conclusion Manufacturer-recommended 5-second UriSponge exposure did not produce clinically significant interference in chemical urinalysis, including SG, pH, or semiquantitative parameters. Prolonged (worst-case) exposure demonstrated acceptable performance overall, though increased pH variability was observed at room temperature, supporting adherence to recommended dipping duration and appropriate storage conditions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A-192 Evaluating the potential impact of leached preservative on chemical urinalysis in residual midstream urine specimens — Science Skim