The British journal of radiology

Filter tilt, dwell time, and IVC diameter in unsuccessful IVC filter retrieval: associations from a single centre experience over ten years.

R Ahmed, R Kernan, N McEniff, J M Ryan, M J Guiney, M O' Donovan, N O' Connell, M Lavin,

Thomas Mitchell Centre for Advanced Medical Imaging, Trinity College Dublin, Dublin 2, D02 PN40, Ireland. · Department of Radiology, St James's Hospital, Dublin 8, D08 NHY1, Ireland.

Observational · n=231

Key points

From the paper's conclusion
  • On multivariable logistic regression, filter tilt at insertion (p = 0.016), dwell time (p = 0.009), and smaller IVC diameter (p = 0.030) were identified as independent risk factors for unsuccessful retrieval.

Abstract

OBJECTIVES: This retrospective cohort study aims to describe management approaches for inferior vena cava (IVC) filter retrieval and identify independent risk factors for unsuccessful retrieval at our tertiary referral centre. METHODS: All patients who underwent IVC filter insertion at our single tertiary referral centre between February 2016 and August 2025 were included. Kaplan-Meier analysis and a multivariable logistic regression were used to determine rates of retrieval and identify independent risk factors for unsuccessful retrieval. A dedicated database for follow-up of patients with IVC filters was managed by the haematology department. RESULTS: A total of 231 patients were included. 103 of 128 (80.47%) retrievals were successful at the first attempt. The retrieval rates at 6 weeks, 3 months, and 1 year were 59.85%, 72.73%, and 81.06%, respectively. On multivariable logistic regression, filter tilt at insertion (p = 0.016), dwell time (p = 0.009), and smaller IVC diameter (p = 0.030) were identified as independent risk factors for unsuccessful retrieval. CONCLUSIONS: Our follow-up database was instrumental in scheduling early removal and minimising loss to follow-up to achieve high retrieval rates. We advocate for the early removal of filters to maximise successful retrieval, avoid the need for advanced removal techniques, and prevent long-term complications. ADVANCES IN KNOWLEDGE: Our findings support hypotheses that prolonged dwell time and greater tilt lead to unsuccessful retrieval and we additionally identify smaller IVC diameter as a previously unreported risk factor. Practically, we provide strong evidence to support the use of a prospectively managed follow-up database for filters.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers