Inferior vena cava filter retrieval with radial jaw biopsy forceps: a case report.
Baylor Akhavan, Brian Kerlin, Rachel Soong, Alexander Kieger, Brandon Chen
Kirk Kerkorian School of Medicine at UNLV, Las Vegas, NV, USA. · Interventional Radiology, University Medical Center of Southern Nevada, Las Vegas, NV, USA.
Key points
From the paper's conclusion- Radial jaw biopsy forceps may serve as a viable alternative in complex IVC filter retrieval when conventional and advanced techniques fail or are unavailable. This case expands the spectrum of available tools and highlights the importance of procedural adaptability in interventional radiology (IR).
Abstract
BACKGROUND: Inferior vena cava (IVC) filters are used in patients at risk for venous thromboembolism (VTE) who cannot receive anticoagulation. Prolonged dwell times are associated with increased risk of embedment, endothelialization, and retrieval failure, often necessitating advanced techniques. While endobronchial forceps are commonly used in complex retrievals, the role of alternative biopsy devices remains limited. We present the following case highlighting the use of radial jaw biopsy forceps for IVC filter retrieval, a technique with minimal prior documentation. CASE DESCRIPTION: A man in his late 30s with a history of recurrent deep vein thromboses presented with progressive bilateral lower extremity swelling and an indwelling IVC filter with an estimated dwell time of approximately 5 years. Imaging demonstrated severe iliocaval obstruction and collateral formation. Initial retrieval attempts using standard snare, loop snare, and hangman techniques were unsuccessful due to fibrotic encapsulation and deformation of the filter. In the absence of rigid endobronchial forceps, radial jaw biopsy forceps were utilized to dissect fibrous tissue and successfully engage and remove the filter. Residual struts were excluded with bilateral iliac vein stenting, resulting in improved venous outflow. CONCLUSIONS: Radial jaw biopsy forceps may serve as a viable alternative in complex IVC filter retrieval when conventional and advanced techniques fail or are unavailable. This case expands the spectrum of available tools and highlights the importance of procedural adaptability in interventional radiology (IR).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Novel endovascular retrieval of duodenal-penetrating inferior vena cava filters: a report of two cases and technical considerations.AME case reports · 2026
- Balloon-assisted radial release for retrieval of a surgically suture-fixed inferior vena cava filter: a case report.Thrombosis journal · 2026
- Cardiac Migration of an Inverted Inferior Vena Cava Filter: Successful Retrieval via a Jugular Loop-Snare Technique.Vascular and endovascular surgery · 2026
- Retrieval of Thrombosed Inferior Vena Cava Filter and Mechanical Thrombectomy of Associated Caval Thrombosis Using FlowTriever and ClotTriever Systems With the ClotTriever as Embolic Protection.Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2026
- Case Report: Clinical application of an adjustable curved sheath in retrieval of difficult-to-retrieve inferior vena cava filters: a retrospective case seriesFrontiers in Cardiovascular Medicine · 2026
- Indwelling inferior vena cava filter as a persistent thrombotic risk factor: recurrent venous thromboembolism after anticoagulation withdrawal.Blood coagulation & fibrinolysis : an international journal in haemostasis and thrombosis · 2026