Inferior Vena Cava Filters in Adolescents with Thromboembolic Disease: Technical Outcomes, Retrieval, and the Role of Skeletal Maturity.
Tomas Melis, Radek Padr, Filip Fencl, David Janak, Lukas Lambert
Department of Imaging Methods, Motol University Hospital and Second Faculty of Medicine, Charles University, Prague, Czech Republic. · Department of Pediatrics, Motol University Hospital and Second Faculty of Medicine, Charles University, Prague, Czech Republic.
Key points
From the paper's conclusion- Off-label IVCF placement in adolescents achieved high technical and retrieval success. Retrieval failure was confined to skeletally immature patients and was not explained by caval diameter. Strut fracture with migration at long-term imaging reinforces the need for timely retrieval.
Abstract
PURPOSE: To evaluate technical success, periprocedural safety and retrieval outcomes of inferior vena cava filter (IVCF) placement in adolescents with venous thromboembolism and to determine whether retrieval failure is associated with skeletal maturity. MATERIALS AND METHODS: Single-center retrospective review of 26 IVCF placed in 25 patients with iliocaval or iliofemoral thrombosis (mean age 15.8 ± 2.2 years; range 9-18) between 2006 and 2024. Skeletal maturity was graded by Risser stage. Retrieval was attempted once the initial indication had resolved and anticoagulation was feasible. RESULTS: Implantation succeeded in every case; mean caval diameter was 17.9 ± 3.0 mm. Retrieval was attempted in 22 of 26 filters and succeeded in 19 (86% of attempts, 79% by intent to retrieve, 73% overall). One successful retrieval required a second attempt with an adjunctive loop-snare maneuver. Median indwelling time was 89 days (range 1-489). All three unsuccessful retrievals occurred at Risser stage 3 or below (3 of 7 versus 0 of 15 attempted; P = .02), not explained by caval diameter (P = .33). Seven filters (27%) remained in situ. No periprocedural adverse events occurred. Both retained filters assessed by CT showed caval wall penetration, one with strut fracture and asymptomatic migration at 16 years; chronic caval occlusion developed in one patient. CONCLUSION: Off-label IVCF placement in adolescents achieved high technical and retrieval success. Retrieval failure was confined to skeletally immature patients and was not explained by caval diameter. Strut fracture with migration at long-term imaging reinforces the need for timely retrieval.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Filter tilt, dwell time, and IVC diameter in unsuccessful IVC filter retrieval: associations from a single centre experience over ten years.The British journal of radiology · 2026
- Multi-Institutional Review of Inferior Vena Cava Filter Use in Pediatric Patients.Journal of vascular and interventional radiology : JVIR · 2026
- Inferior vena cava filter retrieval with radial jaw biopsy forceps: a case report.AME case reports · 2026
- Novel endovascular retrieval of duodenal-penetrating inferior vena cava filters: a report of two cases and technical considerations.AME case reports · 2026
- Case Report: Clinical application of an adjustable curved sheath in retrieval of difficult-to-retrieve inferior vena cava filters: a retrospective case seriesFrontiers in Cardiovascular Medicine · 2026
- Outcomes post-thrombectomy in children: A multi-site assessment of a new non-lytic endovascular device combined with anticoagulation for management of adolescents with extensive acute lower limb deep vein thrombosis.Thrombosis research · 2026