Validation study on cuffless ambulatory blood pressure monitoring method using bio-impedance, impedance cardiography and electrocardiography.
Myrte Schoenmakers, Martin Gevonden, Bruno Sauce, Eco de Geus
Department of Biological Psychology, Vrije Universiteit Amsterdam, Van der Boechorststraat 7, 1081 BT, Amsterdam, The Netherlands. · Department of Biological Psychology, Vrije Universiteit Amsterdam, Van der Boechorststraat 7, 1081 BT, Amsterdam, The Netherlands. eco.de.geus@vu.nl.
Key points
From the paper's conclusion- Although cuffless pulse wave travel time-based BP estimators showed strong reliability, validity was insufficient for clinical or stress research use, even with person-specific calibration. Future work should explore extended calibrationperiods before testing in naturalistic settings.
Abstract
PURPOSE: Continuous, non-invasive blood pressure (BP) monitoring is essential for understanding cardiovascular dynamics in real-world settings. This study evaluated the reliability and validity of a cuffless method for ambulatory BP monitoring using pulse wave travel times derived from bio-impedance, impedance cardiography, and electrocardiography. We examined whether this method can accurately estimate systolic and diastolic BP cross varying conditions. METHODS: Forty-one young, healthy participants completed a protocol involving postural changes, physical activity, and mental stress tasks. Pulse arrival time (PAT) and pulse transit time (PTT) were extracted, and person-specific models estimated BP using (1) all data of the person ('double-dipping') - used in most previous studies -, (2) half of a person's data (calibration), or (3) person-invariant parameters (calibration-free) which would enhance practical utility. Cuff-based BP served as the reference. Test-retest reliability, and construct and convergent validity were assessed. RESULTS: Test-retest reliability was high for PAT- and PTT-based BP for all approaches (ICC > 0.7), comparable to cuff-based measurements. Construct validity was generally poor: BP changes during physical activity were detected, but responses to mental stressors were not. Only double-dipping showed acceptable convergent validity, likely reflecting overfitting. In the calibration approach, convergence with cuff-based BP was moderate (MAE = 10.1-12.9 mmHg), with 59.2-68.5% of estimates within 10 mmHg of reference values. The strongest deterioration of convergentvalidity occurred with person-invariant models. CONCLUSION: Although cuffless pulse wave travel time-based BP estimators showed strong reliability, validity was insufficient for clinical or stress research use, even with person-specific calibration. Future work should explore extended calibrationperiods before testing in naturalistic settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An Unfinished but Necessary Path: Ring-Based Cuffless Blood Pressure Monitoring at the Crossroads of Innovation and Clinical Reality.Pulse (Basel, Switzerland) · 2026
- Continuous Cuffless Blood Pressure Estimation Using PPG From Methods to Clinical TranslationarXiv · 2026
- Evaluation of Alysis-001 cuffless blood pressure estimation algorithm against the European Society of Hypertension awake/asleep test criteria in hypertensive patients.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring.American journal of hypertension · 2026
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring.Clinical hypertension · 2026