Evaluation of GPU-accelerated Monte Carlo beam and agility MLC modeling in the Monaco treatment planning system.

Fasco van Ommen, Sima Sajadi Shahrbabaki, Thomas A Foppen, Jordi Saez, Victor Hernandez, Bram van Asselen, Sara L Hackett

Department of Radiation Oncology, UMC Utrecht, Utrecht, the Netherlands. · Department of Radiation Oncology, Hospital Clínic de Barcelona, Barcelona, Spain.

Key points

From the paper's conclusion
  • Purpose This study evaluates a new GPU‐accelerated Monte Carlo (MC) beam and MLC model (VSM2.0) from Elekta, with a focus on aspects of the MLCs that were previously shown to be particularly challenging to model.

Abstract

BACKGROUND: Treatment planning systems (TPSs) are essential in modern radiotherapy. Accurate beam modeling, especially of multileaf collimators (MLCs), is crucial, as their complex designs often introduce systematic uncertainties in planned and delivered doses. Thorough evaluation of the beam model and MLCs must be performed before clinical introduction. PURPOSE: This study evaluates a new GPU-accelerated Monte Carlo (MC) beam and MLC model (VSM2.0) from Elekta, with a focus on aspects of the MLCs that were previously shown to be particularly challenging to model. METHODS: Measurements were conducted on two Elekta Versa HD linacs with Agility MLCs. All calculations were performed in Monaco v6.2.3 using VSM1.6 and VSM2.0. Open field dose distributions were measured and calculated in a water phantom and compared using a gamma analysis (1%/1 mm gamma criterion). MLC characteristics including the leaf tip penumbra, measured with film, the dosimetric leaf gap (DLG) and the tongue-and-groove effect, as measured via sweeping gap fields, were compared with calculations from both beam models. These measurements also highlighted some challenges associated with MLC model verification. Finally, patient treatment plans with a wide range of target volumes and modulation were measured and compared against calculations (gamma analysis: 3%/2 mm gamma criterion). RESULTS: The VSM2.0 model showed excellent agreement with measurements for open field dose distributions, with gamma pass rates for almost all fields above 90%. The clinical treatment plans showed gamma pass rates above 94%. The VSM1.6 model showed similar accuracy. However, the VSM2.0 modeled leaf transmission more accurately than VSM1.6 and improved representation of the tongue-and-groove effect across different gap widths and offsets. By contrast, the maximum deviations from the measured DLGs were 0.15 mm and 0.43 mm for VSM1.6 and 2.0 respectively, largely due to user-definable tuning of parameters for VSM1.6, which allowed the VSM1.6 model to closely match linac performance. CONCLUSIONS: The new GPU-based single phase-space beam model shows excellent agreement with measurements and improves simulation of complex aspects of the leaf geometry. Reducing customization simplifies the commissioning process of planning systems, thereby potentially reducing associated risks, but the possibility to tune the leaf offset per beam model may further improve the model accuracy.

Read the paper