科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Journal of Instrumentation2026-04-01· Monte Carlo method

Monte Carlo simulation of the ISOLPHARM gamma camera for Ag-111 imaging

Davide Serafini, Aurora Leso, A. Arzenton, S. Spadano, E. Borciani, D. Chen, C. Sbarra, M. Negrini, A. Margotti, Nico Lanconelli, G. Baldazzi, E. Mariotti, S. Corradetti, A. Andrighetto

原始摘要(英文原文)· Original abstract
Abstract Targeted Radionuclide Therapy (TRT) is a well-established technique for cancer treatment. In this approach, radionuclides are bound to specific drugs that selectively transport them to the tumor site. Within the ISOLPHARM project, a radiopharmaceutical for TRT based on the innovative radionuclide 111 Ag is currently under development. 111 Ag has a half-life of 7.45 days and decays by emitting both electrons and gamma-rays. The emission of gamma-rays, predominantly at an energy of 342 keV, enables the visualization of 111 Ag using a gamma camera. In this work, we describe a Monte Carlo simulation developed to optimize the design parameters of such an imaging device. The simulation is based on the Geant4 toolkit, which accurately models the interactions between particles and matter. The estimated spatial resolution and sensitivity of the system are approximately 4 mm and 19 cps/MBq, respectively. The simulated device is able to resolve lesions with a lesion-to-background activity ratio of 4:1 under in-vivo-like conditions. These results indicate that the proposed gamma camera can provide cost-effective imaging capabilities for preclinical radiopharmaceutical studies.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Monte Carlo simulation of the ISOLPHARM gamma camera for Ag-111 imaging — 科研速览 Science Skim