Performance Evaluation of an Arduino‑Driven Medical Ventilator Integrating MAX30100, Thermocouple, and OCS‑3F Sensors for Respiratory Support
Rafiuddin Syam, Aodah Diamah, Daffa Akmal Dewangga
Key points
From the paper's conclusion- The proposed system successfully provides real-time patient monitoring and controlled air-oxygen delivery, offering a viable low-cost alternative for emergency respiratory support in resource-limited settings.
Abstract
This paper presents the development and performance assessment of a low-cost, microcontroller-based emergency ventilator featuring an air-mix intake system. The prototype integrates an Arduino Mega 2560 with a MAX30100 sensor for heart rate and SpO 2 monitoring, a type-K thermocouple for temperature sensing, and an OCS-3F sensor for measuring oxygen purity and flow rate. Additional components include solenoid valves, a mechanical compressor, and a 20 × 4 LCD display. The device operates in three selectable modes: air-only, oxygen-only, and automatic air-oxygen mixing. Experimental results demonstrate that the MAX30100 sensor achieves mean errors of 0.28% for SpO 2 and 0.98% for heart rate when compared to a commercial pulse oximeter. The thermocouple exhibits a temperature reading error ranging from 0% to 2.74%. Over a 60-minute test period, the OCS-3F sensor records an average oxygen purity of 84.43% and an average flow rate of 6.48 L/min, with systematic deviations indicating the need for recalibration. The proposed system successfully provides real-time patient monitoring and controlled air-oxygen delivery, offering a viable low-cost alternative for emergency respiratory support in resource-limited settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Smart oxygen delivery using IoT-based physiological monitoring and adaptive oxygen flow controlDiscover Applied Sciences · 2026
- Development and validation of an integrated wearable system for continuous cardiorespiratory monitoring using photoplethysmography and optical respiratory detection.Medical & biological engineering & computing · 2026
- An IoT-Enabled Multi-Sensor Smart IV Infusion Monitoring System with Edge–Cloud IntegrationApplied System Innovation · 2026
- Simple and easy to build mechanical ventilator for pandemic health crises in resource-limited areas.Frontiers in public health · 2026
- Design and pilot clinical evaluation of a low-cost, continuous respiratory monitor for newbornsmedRxiv · 2026
- A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study.Sensors (Basel, Switzerland) · 2026