An IoT-Enabled Multi-Sensor Smart IV Infusion Monitoring System with Edge–Cloud Integration
Ming-Cheng Kao, Kai-Huang Chen, Kuo-Chien Liao, Fu-Xue Chen, Kuan-Sheng Ho, Yu-Zhen Tong
Key points
From the paper's conclusion- These results demonstrate the feasibility of the integrated architecture at laboratory level but do not establish general measurement accuracy, reliable backflow detection, or clinical accuracy.
Abstract
Conventional gravity-fed intravenous (IV) infusion relies largely on visual inspection, and abnormal events such as incorrect drip rate, depletion, and blood backflow may occur between nursing rounds. This study develops and evaluates an IoT-enabled multi-sensor bedside monitoring system with edge–cloud integration. A dual-core ESP32 edge node integrates SD01 optical drip-rate sensing, XKC-Y25 non-contact depletion detection, TCS34725 optical blood-backflow detection, MAX30105 heart-rate and SpO2 monitoring, and PN532 NFC/RFID staff authentication. FreeRTOS task partitioning separates high-rate physiological processing from communication tasks. Deterministic alarms are generated locally on the edge node, independent of cloud availability. Processed states are synchronized to a Supabase cloud backend every 1 s and displayed on a multi-bed dashboard with privacy-aware public and NFC-authenticated clinical views. Laboratory tests showed an aggregate IV flow-rate mean absolute error below 3.5% and a backflow alarm response time below 1 s; the sensitivity, specificity, and false-positive rate of the backflow detector were not quantified. In a preliminary comparison with a commercial pulse oximeter (five paired readings), the mean differences were 4.2 bpm for heart rate and 0.6 percentage points for SpO2, with mean absolute differences of 5.0 bpm and 1.8 percentage points, respectively. These results demonstrate the feasibility of the integrated architecture at laboratory level but do not establish general measurement accuracy, reliable backflow detection, or clinical accuracy. A knowledge-guided decision-support layer is outlined as future work.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Smart oxygen delivery using IoT-based physiological monitoring and adaptive oxygen flow controlDiscover Applied Sciences · 2026
- ESP32-S3-Based Multimodal Wearable for Remote Vital-Sign Acquisition with ECG-to-PPG Pulse-Arrival-Time Tracking and Functional Cloud Triage: An Engineering Proof of Concept.Sensors (Basel, Switzerland) · 2026
- Edge-Native Non-Contact Heart Rate Monitoring via Wi-Fi CSI on Low-Cost MicrocontrollersACM Transactions on Embedded Computing Systems · 2026
- HL7 FHIR-Based Open-Source Framework for Real-Time Biomedical Signal Acquisition and IoMT InteroperabilityApplied Sciences · 2025
- An Edge–Fog–Cloud IoT Framework for Real-Time Cardiac Monitoring and Rapid Clinical Alerts in Hospital WardsFuture Internet · 2026
- Physical and Synthetic Evaluation of Event Integrity, BLE Transmission, and Browser Reconstruction in the PHUKUY SpirometerApplied Sciences · 2026