Journal of Physics: Conference Series

Development of double-clad Qdot-based microstrip detector for Spatially Fractionated Radiotherapy dosimetry

Josue Esteves, Pierre Everaere, Thu Nhi Tran Caliste, Benjamin Fornacciari, Patrick Pittet

Key points

From the paper's conclusion
  • To address the spatial resolution challenge posed by dosimetry in spatially fractionated radiotherapy, we have developed a dual-clad microstrip dosimetry detector based on CdSe/ZnS quantum dots.

Abstract

To address the spatial resolution challenge posed by dosimetry in spatially fractionated radiotherapy, we have developed a dual-clad microstrip dosimetry detector based on CdSe/ZnS quantum dots. The spatial resolution of 20µm of this detector was measured using a 10 µm wide X-ray microbeam at the ESRF and reaches the current state of the art.