Characterization study of 20 µm and 60 µm wide BC-408 microstrips during microbeam X-ray irradiation

F. Thevenet, J. Esteves, J-P. O Bustillo, E. Li, M. J. Lerch, J-F. Adam, P. Pittet

Key points

From the paper's conclusion
  • We also observed a relatively stable response of the strips as a function of cumulative dose (with a deviation of less than 4% for cumulative doses up to 30 kGy).

Abstract

A radiation detector implementing 20 µm and 60 µm wide BC-408 microstrips has been developed and characterized during microbeam X-ray irradiations at very high dose rates. The 20-µm-wide strips allow for a fairly accurate measurement of the microbeam profiles, with a slight broadening of their base, but with a lower scintillation efficiency than that of the 60-µm-wide strips. This decrease in efficiency has been attributed to the two-dopant scintillation mechanism in BC-408, suggesting that, for 20 µm-wide strips, a scintillator material with higher concentrations of primary fluorophores and wavelength shifter fluors should be used. We also observed a relatively stable response of the strips as a function of cumulative dose (with a deviation of less than 4% for cumulative doses up to 30 kGy). The BC-408 strips exhibited a marked energy dependence for the tested beam energies ( HVL Cu ∈ [0.27 mm ; 2.28 mm ]) which needs to be taken into account for MRT dosimetry applications.