Cherenkov-excited fluorescence imaging for small-field dosimetry and SRS plan verification: a feasibility study.
Mehmet Fazil Enkavi, Hasan Uysal, Kaan Oysul, Sait Sirin, Hülya Yavuz Ersan
Bioengineering, Hacettepe University, Hacettepe University, Department of Bioengineering, Beytepe, Çankaya, Ankara, 06800, Turkey. · Radiation Oncology, Medicana International Ankara Hospital, Söğütözü, 2176. Sk. No: 3, Çankaya, Ankara, 06510, Turkey.
Key points
From the paper's conclusion- This study demonstrates compact, high-spatial-resolution, rapid-readout optical dosimetry for small-field and patient-derived SRS dose patterns.
Abstract
To develop a high-spatial-resolution Cherenkov-excited fluorescence dosimetry system (CEFD) with rapid optical readout using quinine sulfate solution and evaluate its feasibility for small-field dosimetry and verification of patient-derived stereotactic radiosurgery (SRS) dose distributions. Approach. A 3D-printed optical phantom containing 1 g/L quinine sulfate solution was imaged with a CMOS camera through a 45° mirror. Sensitive-volume thicknesses of 2.5, 5, 10 and 15 mm were compared using calibration response, signal-to-noise ratio, output factors, dose-rate dependence, and small-field beam profiles benchmarked against a PTW 60023 microSilicon detector. The selected CEFD configuration was further evaluated for thermal response, repeatability, signal drift, low-MU response, and small-field gamma performance, and then used for fixed-gantry verification of 31 patient-derived SRS dose distributions. GafChromic EBT3 film was used as the high-spatial-resolution comparator. Main results. The CEFD achieved 0.065 mm/pixel spatial sampling. CEFD-10 showed the most balanced overall performance across sensitivity, dose-rate response, output-factor agreement, and small-field profile measurements. Dose-rate deviations remained within 1.7% at 300-1400 MU/min. Passive heat sinking reduced the coefficient of variation from 0.26% to 0.13% and limited signal drift from 4.3% to 0.97% over 30 consecutive measurements. For the 0.5 × 0.5 cm² field, the CEFD-10 central-axis dose ratio relative to the TPS-projected dose remained within ±2% from 10 MU onward. For 0.5 × 0.5-3 × 3 cm² static fields, CEFD-10 achieved 99.56%-100.0% global 2%/1 mm gamma passing rates with a 10% low-dose threshold. For patient-derived SRS verification, CEFD-10 achieved mean global gamma passing rates of 99.47 ± 1.58% and 97.53 ± 3.58% for 2%/2 mm and 2%/1 mm criteria, respectively, with mean differences from EBT3 film of -0.32% and -0.21%, respectively. Significance. This study demonstrates compact, high-spatial-resolution, rapid-readout optical dosimetry for small-field and patient-derived SRS dose patterns. Agreement with EBT3 film supports CEFD-10 as a controlled-geometry SRS dose verification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative Cherenkov-excited fluorescence emission dosimetry for HDR Co-60 brachytherapy: Calibration and planar validation against TPS.Journal of applied clinical medical physics · 2026
- From Radiotherapy Dataset to Cherenkov Dose Map: A Digital Linac Monte Carlo Simulation PipelinearXiv · 2026
- Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR BrachytherapyarXiv · 2026
- Dosimetric Characterization and Workflow Optimization of the FLASH-SARRP for Reliable Preclinical Radiobiological StudiesbioRxiv · 2026
- Detective quantum efficiency (DQE)-oriented design of a transparent scintillator tandem x-ray/optical microscope for photon-starved 7.2-keV radiography at the National Ignition Facility (NIF).The Review of scientific instruments · 2026
- Experimental Validation and Calibration of Cherenkov Imaging for Real-Time Surface Dosimetry in Photon Breast Therapy.Proceedings of SPIE--the International Society for Optical Engineering · 2026