Optical measurement and calibration methods of coagulation analyzer
Yang Liu, Bin Liu, Fei Song, Haiyan Mo, Ziyan Zhao, Hairong Yu, Yannan Ren
Key points
From the paper's conclusion- Metrological calibration of these optical systems was rigorously conducted in accordance with JJF 1945-2021 “Calibration Specification for Coagulation Analyzers.” Major performance indicators, including the optical response to fibrinogen (FIB), prothrombin time (PT), activated partial thromboplastin time (APTT), and thrombin time (TT), were quantitatively analyzed.
Abstract
Optical measurement is the fundamental mechanism for modern coagulation analyzers, typically relying on turbidimetric variations to detect clot formation. Metrological calibration of these optical systems was rigorously conducted in accordance with JJF 1945-2021 “Calibration Specification for Coagulation Analyzers.” Major performance indicators, including the optical response to fibrinogen (FIB), prothrombin time (PT), activated partial thromboplastin time (APTT), and thrombin time (TT), were quantitatively analyzed. Key metrological characteristics, such as indication error and repeatability, were investigated to assess the stability of the optical signal acquisition. Based on the principle of biosensing, this study found that fluctuations in incubation temperature significantly affect thrombin activity, thereby causing deviations in optical density signals during FIB measurements. Consequently, a refined metrological traceability system was established to ensure the reliability of the photonic sensing process in clinical laboratories. This work provides a comprehensive reference for the optical performance verification and calibration of coagulation analyzers.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-231 Tri-Modal Detection in Coagulation Testing: Performance Evaluation of the Zybio EXT 9800 Coagulation Analyzer and Its Utility in Challenging Clinical SamplesClinical Chemistry · 2026
- Numerical and Interpretive Comparability of Optical and Mechanical Coagulation Analyzers: A Zlog-Supported Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026
- A-245 Robust Performance of the Quantra® Hemostasis Analyzer in Global Proficiency TestingClinical Chemistry · 2026
- From Signals to Measurands: A Measurement-Science Roadmap for Reproducible Analytical Biochemistry.ACS measurement science au · 2026
- Calibration Errors and their Impact on Measurement Accuracy in Lifetime based Phosphor ThermometryMeasurement Science and Technology · 2026
- Analytical performance comparison of Siemens Atellica CI1900 and Roche Cobas Pure integrated analyzers across spectrophotometric, ion-selective electrode and immunoassay parameters.Analytical biochemistry · 2026