Spot-wise compensation of proton beam deflection and radiation field deformation in the transverse magnetic field of an in-beam MR scanner using a research version of a commercial treatment planning system.
Marisa Cobanaj, Erik Traneus, Felix Horst, Krishna Jeanne Godino Padre, Stefan Menkel, Petra Trnkova, Aswin L Hoffmann, Franciska Lebbink
Faculty of Medicine, and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, OncoRay - National Center for Radiation Research in Oncology, Händelallee 26, Bldg. 130, Dresden, 01309, Germany. · RaySearch Laboratories AB, Sveavagen 25, SE-11 34 Stockholm, Stockholm, Stockholm County, 104 30, Sweden.
Key points
From the paper's conclusion- This work aims to develop and experimentally validate a spot-wise compensation method capable of restoring proton dose distributions in the presence of a measured 3D magnetic vector field map within a clinical treatment planning system (TPS).
Approach
A spot-wise compensation framework was implemented in a research version of RayStation, a TPS commercially available for clinical use.
Abstract
Objective Magnetic resonance-integrated proton therapy (MRiPT) offers the potential for high-precision, image-guided particle therapy but is challenged by magnetic field (MF)-induced proton beam deflection and radiation field deformation. Unless compensated for, these effects compromise dose conformity and prohibit clinical implementation. This work aims to develop and experimentally validate a spot-wise compensation method capable of restoring proton dose distributions in the presence of a measured 3D magnetic vector field map within a clinical treatment planning system (TPS). Approach A spot-wise compensation framework was implemented in a research version of RayStation, a TPS commercially available for clinical use. The method compensates both lateral beam deflection and MF-induced radiation field deformation by shifting individual pencil-beam spots based on displacement vectors derived from dose calculations in water phantoms using the measured MF of an in-beam MR scanner integrated with a horizontal proton beamline. Two complementary strategies were implemented: a patient-specific approach based on plan-specific displacement calculations, and a generic approach using precomputed displacement maps derived from a reference phantom. Compensation performance was evaluated in representative clinical cases. Dose distributions were experimentally validated using 2D ionisation chamber array measurements. Main Results Both compensation methods effectively restored target coverage, dose homogeneity, and organ-at-risk sparing to the level of no-MF reference plans. The patient-specific method generally achieved a higher performance, whereas the generic compensation method enabled rapid automated correction with sufficient accuracy for clinical workflows involving deep-seated tumours. Local γ-analysis demonstrated excellent agreement between measured and calculated doses, with pass rates exceeding 98.5% for 3%/3 mm criteria. Significance This work demonstrates that compensation of MF-induced proton beam deflection and field deformation can be achieved entirely within a research version of the RayStation TPS using measured, MR scanner-specific MF data. The proposed method provides a scalable, robust framework for future MRiPT clinical applications.
.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and Experimental Validation of an Energy Layer-wise Beam Deflection Compensation Method for MR-integrated Proton Therapy with a Transverse Magnetic Field.International journal of radiation oncology, biology, physics · 2026
- Dosimetric validation of proton beam deflection and field shape deformation in a measured transverse magnetic field of an in-beam MR scanner using a novel dose calculation functionality in a research version of a commercial treatment planning system.Physics in medicine and biology · 2026
- Magnetic-Field-Induced Secondary-Electron Effects in MR-Guided Radiotherapy: From Dosimetric Mechanisms to Workflow-Based Clinical Risk Management.Technology in cancer research & treatment · 2026
- Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgeryJournal of Applied Clinical Medical Physics · 2026
- Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy systemarXiv · 2026
- Magnetically enhanced peak‐to‐valley dose ratio for megavoltage photon minibeam radiotherapyMedical Physics · 2026