In silico biomechanical evaluation of personalized A-shaped titanium plates for mandibular condylar base fractures.
Jie Li, Chong-Tao Xu, Ting Luo, Guo-Ning Xiao, Yuan Liang
Department of Stomatology, Tianjin Medical University General Hospital, Tianjin, PR China.
Key points
From the paper's conclusion- This study developed personalized A-shaped plates with 5-, 6-, and 7-hole configurations, geometrically optimized to adjust the curvature of the anterior arm to accommodate the anatomical contour of the sigmoid notch.
Abstract
The anatomical complexity of the mandibular condylar base renders open reduction and internal fixation challenging for fractures in this region. Although the double-straight-plate fixation remains the most widely adopted clinical strategy, its implementation is frequently hindered by technical limitations inherent to conventional operative routes. The biomechanical adequacy of single-geometric-plate fixation has not been conclusively established and warrants rigorous experimental validation. This study developed personalized A-shaped plates with 5-, 6-, and 7-hole configurations, geometrically optimized to adjust the curvature of the anterior arm to accommodate the anatomical contour of the sigmoid notch. Subsequently, the biomechanical efficacy of these personalized plates in stabilizing condylar base fractures was systematically evaluated. Finally, a multi-dimensional quantitative comparison was conducted to assess biomechanical differences between the proposed single-geometric-plate fixation and conventional fixation protocols. The 7-hole-A-shaped-plate configuration demonstrated biomechanical performance equivalent to the double-straight-plate configuration across all parameters. The 6- and 5-hole-A-shaped-plate configuration achieved comparable global stability but exhibited significantly elevated stress concentration in either fixation hardware or adjacent cortical bone. The in silico study furnishes robust biomechanical evidence to support individualized selection of A-shaped geometric fixation for condylar base fractures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- PEEK plates for the fixation of mandibular condylar base fractures: a finite element analysis comparison of single and double plate configurations with titanium and resorbable systems.Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · 2026
- Influence of Plate Design, Thickness, and Fixation Architecture on Mandibular Advancement Stability: A Finite Element AnalysisJournal of Clinical Medicine · 2026
- Subject-Specific Finite Element Modelling of Le Fort I Osteotomy Fixation: Biomechanical Insights Comparing Patient-Specific Implants and Conventional Miniplates.Journal of clinical medicine · 2026
- Finite element analysis of patient specific three-dimensional titanium plate versus conventional two miniplate fixation for mandibular angle fractures.BMC oral health · 2026
- Biomechanical Evaluation of Patient-Specific Titanium Fixation Systems for Segmental Le Fort I Osteotomy: A Finite Element Analysis.Journal of functional biomaterials · 2026
- Biofiligree®—A Concept for Bone Fracture Fixation PlatesSci · 2026