Comparison of refractive error measurements obtained with a pyramid wavefront sensor topo-aberrometer and an auto kerato-refractometer against subjective refraction.
Alex Julien, Gemma Julio, Fathima Junaidha, Rafael I Barraquer
Optometry Department, Barraquer Eye Hospital, Dubai, United Arab Emirates. · Research Department, Centro de Oftalmología Barraquer, Barcelona, Spain.
Key points
From the paper's conclusion- Both devices provide a useful reference, but they are not interchangeable with SR. KR-800 demonstrated closer agreement with SR across all parameters.
Abstract
CLINICAL RELEVANCE: Refraction is a crucial part of an ocular examination. Ensuring the accuracy and efficiency of objective baseline measurements is essential for optimising clinical workflow and improving final prescription outcomes. BACKGROUND: Agreement between objective refraction of Topcon KR-800 auto kerato-refractometer and CSO Osiris-T topo-aberrometer with subjective refraction (SR), has never been evaluated before. This study aimed to evaluate the agreement and clinical interchangeability of these devices with SR. METHODS: This observational cross-sectional study included patients with best corrected distance visual acuity ≥6/7.5. The proportions of spherical equivalent (SE) differences within ±0.25 D and ±0.50 D were reported descriptively. SE, cylinder, axis, J0 and J45 were compared, using Bland-Altman plots with 95% limits of agreement, and double-angle astigmatic plots of vectorial cylinder difference. RESULTS: The sample included 118 eyes of 60 women and 58 men with a mean (SD) age of 35.6 years (11.4), range 18 to 64 years. The percentages of differences between KR-800 and Osiris-T with SR were, respectively, 68.97% and 29.31% within ±0.25 D, and 92.24% and 59.48% within ±0.50 D. The mean spherical equivalent difference with SR was -0.09 D for KR-800 and 0.33 D for Osiris-T, and the 95% limits of agreement were -0.68 to 0.51 D and -0.54 to 1.21 D, respectively. The centroid of the vectorial astigmatic differences with SR was 0.04 D at 72º for KR-800 and 0.24 D at 41º for Osiris-T. Mean axis difference with SR was 2.47º (95% limits of agreement = -10.28 to 15.23º) for KR-800, and -2.73º (95% limits of agreement = -33.22 to 27.76º) for Osiris-T. CONCLUSION: Both devices provide a useful reference, but they are not interchangeable with SR. KR-800 demonstrated closer agreement with SR across all parameters.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact and Limitations of Front-Surface Reflection-Based and Anterior Segment OCT-Derived Keratometry on Refractive Prediction Accuracy in Intraocular Lens Power Calculation.Journal of clinical medicine · 2026
- Agreement of keratometry measurements for toric IOL planning in routine clinical practice: KR-800, IOLMaster 500, and Pentacam HR.PloS one · 2026
- Comparison of Refraction Obtained with the Myopia Master and Subjective Manifest Refraction in Pediatric Orthokeratology.Diagnostics (Basel, Switzerland) · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Repeatability and Cross-Domain Agreement of Pyramidal-Sensor Ocular Wavefront Measurements in Myopic Eyes: A Registry Validation of OSIRIS.Translational vision science & technology · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026