Agreement of keratometry measurements for toric IOL planning in routine clinical practice: KR-800, IOLMaster 500, and Pentacam HR.
Yeoeun Lim, Moonwon Hwang
Cognitive Science Program, College of Arts and Sciences, Indiana University, Bloomington, Indiana, United States of America. · Department of Ophthalmology, Inje University Busan Paik Hospital, Inje University College of Medicine, Busan, Republic of Korea.
Key points
From the paper's conclusion- This retrospective study evaluated keratometry agreement among the KR-800, IOLMaster 500, and Pentacam HR under real-world clinical conditions in 227 eyes of cataract surgery candidates.
Abstract
Accurate keratometry is essential for toric intraocular lens (IOL) planning, yet most prior studies evaluated inter-device agreement under standardized research conditions that may not reflect everyday practice. This retrospective study evaluated keratometry agreement among the KR-800, IOLMaster 500, and Pentacam HR under real-world clinical conditions in 227 eyes of cataract surgery candidates. Corneal astigmatism was converted to power vector components: J0 (with/against-the-rule astigmatism) and J45 (oblique astigmatism). Population-level agreement was assessed using linear mixed-effects models and intraclass correlation coefficients (ICC), while individual-level agreement was evaluated using Bland-Altman analysis. Agreement in toric IOL candidacy and the predicted residual astigmatism arising from inter-device differences were also examined. No significant population-level differences were observed among devices (J0: p = 0.138; J45: p = 0.233), with ICC values of 0.894 for J0 and 0.748 for J45, respectively. However, the 95% limits of agreement were wide (0.80-1.21 D). When any device pair was considered, the predicted residual astigmatism exceeded 0.50 D in 64.8% of patients, and the Pentacam HR classified fewer eyes as toric IOL candidates than the reflection-based devices. Among toric IOL candidates (≥1.0 D astigmatism), 21-33% demonstrated axis differences exceeding 10°, a threshold associated with clinically relevant reduction in astigmatic correction. Axis differences were inversely correlated with astigmatism magnitude (Spearman's rho = -0.220 to -0.272, all p-values < .001). The three devices showed acceptable agreement at the population level, and notably, the average magnitude of disagreement was limited. At the individual level, however, differences exceeded clinically relevant thresholds in a meaningful proportion of patients, including some toric IOL candidates. Using a single device consistently throughout surgical planning may be prudent, although the clinical impact of these differences was not directly assessed and warrants prospective validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simulated Keratometry Versus Total Keratometry: Separate Evaluations in 932 Eyes with Prior Myopic LASIK and 372 Eyes with Prior Radial Keratotomy.Journal of clinical medicine · 2026
- Early Visual Outcomes of Toric Versus Non-Toric IOLs in Low-to-Moderate Corneal Astigmatism: A Real-World Matched Cohort Study.Journal of clinical medicine · 2026
- Use of the internal astigmatism axis to determine the toric intraocular lens axis: Comparison between C-loop and plate-haptic lenses.Arquivos brasileiros de oftalmologia · 2026
- Impact and Limitations of Front-Surface Reflection-Based and Anterior Segment OCT-Derived Keratometry on Refractive Prediction Accuracy in Intraocular Lens Power Calculation.Journal of clinical medicine · 2026
- Comparison of refractive error measurements obtained with a pyramid wavefront sensor topo-aberrometer and an auto kerato-refractometer against subjective refraction.Clinical & experimental optometry · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026