Agreement between lunar densitometers in the capital region of Denmark-a phantom study.
Marcus Bech Edelbo, Bo Zerahn
Department of Clinical Physiology and Nuclear Medicine, Copenhagen University Hospital Herlev, Denmark.
Key points
From the paper's conclusion- This study evaluated the agreement and equivalence between nine Lunar iDXA and four Lunar Prodigy scanners, assessing their interchangeability for longitudinal spine BMD monitoring.
Abstract
In clinical practice, patient follow-up requires scans to be performed on the same densitometer to ensure comparability. This study evaluated the agreement and equivalence between nine Lunar iDXA and four Lunar Prodigy scanners, assessing their interchangeability for longitudinal spine BMD monitoring. The manufacturer's encapsulated spine phantom was scanned 15 times on each of nine iDXA and four Prodigy densitometers under routine quality-control conditions by a single experienced operator. For each scanner, L2-L4 BMD repeatability was assessed using mean, SD, and coefficient of variation. Between-scanner equivalence was tested pairwise using two one-sided tests with ±1% equivalence bounds with Holm-Bonferroni correction. Exploratory method-comparison plots were used to assess agreement across vertebral regions. All iDXA scanners demonstrated low within-scanner variability (CV: 0.07-0.23%) and closely clustered mean BMD values, with a small negative bias relative to the phantom nominal value (mean -0.63%). All pairwise iDXA comparisons met the predefined ±1% equivalence criterion. Graphical analyses confirmed high agreement across all vertebral levels. In contrast, Prodigy scanners showed higher variability (CV: 0.17-0.33%), a positive bias (mean +1.5%), and failed to meet equivalence criteria in most pairwise comparisons. Under the conditions of this phantom study, the nine iDXA scanners in the region showed high agreement and may be sufficiently equivalent for longitudinal spine BMD monitoring across different sites. However, the Prodigy scanners did not demonstrate sufficient agreement for interchangeable use. These findings support regional harmonization of iDXA-based BMD measurements, although validation in vivo, including femoral and full-body measurements, is warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers.Archives of osteoporosis · 2026
- Age-related bone mineral density patterns in a large German clinical DXA cohort: device-specific comparisons with international reference populations.Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA · 2026
- Utility of photon-counting CT localizer radiograph in opportunistic bone densitometry.Journal of medical imaging (Bellingham, Wash.) · 2026
- 3D-DXA cortical and trabecular parameters: agreement between hologic densitometers in clinical practice.Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry · 2026
- Site-Specific Material Decomposition Versus Universal Z-Effective: Optimizing Opportunistic Bone Mineral Density Screening via Dual-Energy CT.Journal of imaging · 2026
- CT Radiomics and Machine Learning for Three-Class Bone Density Classification: An Exploratory Comparison Across Lumbar Vertebral Levels.Tomography (Ann Arbor, Mich.) · 2026