3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers.
Diane Krueger, Neil Binkley, Miguel Madeira, Zhaojing Chen, Silvana Di Gregorio, Luis Del Río, Marta Bracco, Ludovic Humbert
Osteoporosis Clinical Research Program and School of Medicine and Public Health, University of Wisconsin, Madison, WI, USA. dckruege@wisc.edu. · Osteoporosis Clinical Research Program and School of Medicine and Public Health, University of Wisconsin, Madison, WI, USA.
Key points
From the paper's conclusion- Under the conditions evaluated, aBMD and 3D-DXA bone measurements demonstrate comparable short-term precision and inter-scanner diagnostic classification agreement across Prodigy and iDXA densitometers.
Abstract
UNLABELLED: We assessed agreement and precision of 3D-DXA parameters between Prodigy and iDXA densitometers in adults. Strong agreement was observed between devices, and short-term precision was comparable. Moreover, high inter-scanner diagnostic classification agreement was found in the cohort spanning the full range of bone density categories, supporting the use of 3D-DXA across these GE Healthcare models for cross-sectional diagnostic purposes. PURPOSE: 3D-DXA reconstructs DXA hip scans to 3-dimensional images, allowing measurement of trabecular and cortical bone parameters. Differences in imaging characteristics between the GE Healthcare iDXA and Prodigy could influence 3D-DXA results. This study evaluated the agreement and precision of 3D-DXA cortical and trabecular femur parameters between Prodigy and iDXA densitometers. METHODS: A total of 391 adults from 3 clinical centers in the United States and Brazil underwent hip scans on both Prodigy and iDXA scanners, subsequently analyzed with 3D-Shaper software. Short-term precision was assessed on three Prodigy and two iDXA densitometers. Agreement was assessed by regression, Bland-Altman analyses, and diagnostic classification agreement analysis. Short-term precision was determined following International Society for Clinical Densitometry recommendations. RESULTS: Bland-Altman analyses demonstrated statistically significant differences in aBMD and 3D-DXA measurements between Prodigy and iDXA scanners; however, diagnostic classification agreement was high. In the largest cohort, the only one spanning all three classification categories, agreement was 95.5% for cortical sBMD and 95.2% for trabecular vBMD (κw = 0.90 for both), comparable to total hip aBMD (97.3%, κw = 0.94). Short-term precision errors (RMS-SD) were 0.006 to 0.011 g/cm² for total hip aBMD, 0.009 to 0.023 g/cm² for femoral neck aBMD, 2.989 to 4.139 mg/cm³ for integral vBMD, 2.116 to 2.392 mg/cm² for cortical sBMD and 3.073 to 3.909 mg/cm³ for trabecular vBMD. CONCLUSION: Under the conditions evaluated, aBMD and 3D-DXA bone measurements demonstrate comparable short-term precision and inter-scanner diagnostic classification agreement across Prodigy and iDXA densitometers.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 3D-DXA cortical and trabecular parameters: agreement between hologic densitometers in clinical practice.Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry · 2026
- Age-related reference data for cortical and trabecular 3D-DXA parameters: the SEIOMM-3D-DXA project.Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry · 2026
- Comparison of 3D-DXA Measurements with QCT Software Solutions for Proximal Femur Bone Density AssessmentarXiv · 2026
- Agreement between lunar densitometers in the capital region of Denmark-a phantom study.Scandinavian journal of clinical and laboratory investigation · 2026
- Age-related bone mineral density patterns in a large German clinical DXA cohort: device-specific comparisons with international reference populations.Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA · 2026
- Cortical and trabecular deficits in patients with recent hip fractures assessed by DXA and 3D-DXA.Journal of bone and mineral metabolism · 2026