3D-DXA cortical and trabecular parameters: agreement between hologic densitometers in clinical practice.
Marta I Bracco, Jorge Malouf, Laurent Maïmoun, Xavier Nogues, Jean Paul Roux, François DuBoeuf, Ludovic Humbert
3D-Shaper Medical, Barcelona, Spain. · Mineral Metabolism Unit, Grup Creu Groga, Mataró, Spain.
Key points
From the paper's conclusion- Standard and three-dimensional hip bone parameters showed excellent agreement across the evaluated Hologic scanners, with inter-scanner differences rarely affecting density classification. Further studies are needed to confirm generalizability to other Hologic devices.
Abstract
BACKGROUND: Three-dimensional dual-energy X-ray absorptiometry (DXA) reconstructs three-dimensional maps of the proximal femur's density distribution from standard hip DXA scans, enabling the estimation of trabecular and cortical bone parameters. The aim of this study was to assess the agreement of these three-dimensional cortical and trabecular femur parameters across different series and models of Hologic densitometers. METHODOLOGY: The study cohort was composed of 103 women and men recruited from four clinical centers in Spain and France. Subjects had duplicated hip DXA scans using different Hologic scanners from the Horizon, Discovery, and QDR 4500 series. Analyses were performed using 3D-Shaper software. Inter-scanner agreement was evaluated using Deming regression and Bland-Altman analysis. A multiple linear regression was used to investigate the impact of patients' age and BMI on the magnitude of inter-scanner differences. Furthermore, the clinical impact of inter-scanner variability was assessed by evaluating the agreement in the classification of DXA and 3D-Shaper bone parameters as normal, low, or very low density. RESULTS: The parameters demonstrated strong inter-device agreement across all clinical centers and scanner models, with coefficients of determination greater than 0.91. Absolute biases were less than 2.5 mg/cm3 for integral volumetric (v) bone mineral density (BMD), less than 2.9 mg/cm3 for trabecular vBMD, and less than 1.7 mg/cm2 for cortical surface (s) BMD. No statistically significant bias was found between parameters obtained from different scanners. Furthermore, a high classification agreement was found for all parameters, with cortical sBMD reaching the agreement level of standard DXA areal (a) BMD. CONCLUSIONS: Standard and three-dimensional hip bone parameters showed excellent agreement across the evaluated Hologic scanners, with inter-scanner differences rarely affecting density classification. Further studies are needed to confirm generalizability to other Hologic devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers.Archives of osteoporosis · 2026
- Age-related reference data for cortical and trabecular 3D-DXA parameters: the SEIOMM-3D-DXA project.Journal of clinical densitometry : the official journal of the International Society for Clinical Densitometry · 2026
- Comparison of 3D-DXA Measurements with QCT Software Solutions for Proximal Femur Bone Density AssessmentarXiv · 2026
- Cortical and trabecular deficits in patients with recent hip fractures assessed by DXA and 3D-DXA.Journal of bone and mineral metabolism · 2026
- Age-related bone mineral density patterns in a large German clinical DXA cohort: device-specific comparisons with international reference populations.Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA · 2026
- Agreement between lunar densitometers in the capital region of Denmark-a phantom study.Scandinavian journal of clinical and laboratory investigation · 2026